Cooltech equipment consists of 3 large blocks: an electronic block - for control of all inputs and
outputs of the equipment, a refrigeration block
– for cooling the hot part of the peltiers and a
vacuum block - for stopping blood circulation in the treated area, avoiding heating the area for
better results. Each of these blocks is described in more details below.
Electronics:
The electronic block consists of 2 CPUs: the Main and TFT control board.
Main CPU
TFT control board
Main CPU is located inside the electic cabinet of the equipment (on the left side, viewed from the
front), while TFT control board is mounted underneath the display.
TFT control board is responsible for displaying the necessary images on the screen, analyzing
and informing the Main CPU, when the user enters any data and commands on TFT.
All images, graphics and active zones for the screen are stored on a microSD inserted in TFT
control board.
Summary of Contents for Cooltech
Page 1: ...Cooltech SERVICE MANUAL Service Manual Cooltech V1 1 ...
Page 5: ... A4 01454 x2 A4 01453 A4 01454 A4 01455 A4 01455 A4 01455 x2 x2 x2 ...
Page 7: ...Packing layout A4 02266 A4 01389 x2 A4 01591 A4 01391 A4 01415 ...
Page 8: ......
Page 12: ...1 1 2 Vacuum system ...
Page 37: ...Power supply mounting screws For better access to the screws ...
Page 50: ...Power supply screws For better access to the screws ...