3.1.5 -
Vacuum circuit cleaning reminder
Vacuum circuit cleaning warning, appears every 100 treatments and is a reminder only. If turn on the
machine or return to home screen 100 times, this reminder will reappear.
The equipment does not have any device to know that the cleaning has been carried out, so it
is not recommended to reset the warning only.
Note: Once at least a 5 minute treatment is performed, the warning will disappear.
3.1.6 - Applicator disconnected
This alert can appear on both the treatment and main screen, when the accessory recognition system
stops reading that the handle is connected.
This error occurs normally if the applicator is not well connected, so to ensure a good contact check
that the metal bar is totally locking the front connector. If it is closed properly and the alert reappear
intermittently, access the code screen and check if the reading value is stable. Enter the code 65001
for the applicator 1 and the 65002 for the applicator 2. Move the applicator hose and if the value is
not stable proceed to repair the applicator`s connector as it might have a false contact.
Different resistance value are installed, depending on the type of applicator:
Summary of Contents for Cooltech
Page 1: ...Cooltech SERVICE MANUAL Service Manual Cooltech V1 1 ...
Page 5: ... A4 01454 x2 A4 01453 A4 01454 A4 01455 A4 01455 A4 01455 x2 x2 x2 ...
Page 7: ...Packing layout A4 02266 A4 01389 x2 A4 01591 A4 01391 A4 01415 ...
Page 8: ......
Page 12: ...1 1 2 Vacuum system ...
Page 37: ...Power supply mounting screws For better access to the screws ...
Page 50: ...Power supply screws For better access to the screws ...