3.2.2 - Accessories not recognized
When accessing to the pre-treatment screen, the following image appears in any of the channels:
Applicator not connected or resistance value not correct
Indicates either that there is no applicator connected or that the measured resistance value is not
correct. Check first if the applicator is properly connected (the metal bar is locking completely the
front connector). The next step, in case the problem persists, is to check the resistance value the
equipment is reading.
To do this, access the code screen and type codes 65001 for applicator 1 and 65002 for applicator 2.
The corresponding values for each type of applicators are as shown in below. Any other value out of
these ranges, the equipment will interpret it as that there is no accessory connected.
Applicator type
Resistance value
Machine accepting range
Straight
0 R
0-40
Tight
.Curved
Oval
7,68Kohm
592-645
Tiny
4,22Kohm
133-181
Double
6,04Kohm
440-492
No applicator
-
1023-1024
In case of values 1023-1024, it is recommended to test the same applicator on the other channel to
exclude possible failure of the applicator. If the value is still the same check the recognition wiring
connected to the Main CPU (marked in red on the image).
Summary of Contents for Cooltech
Page 1: ...Cooltech SERVICE MANUAL Service Manual Cooltech V1 1 ...
Page 5: ... A4 01454 x2 A4 01453 A4 01454 A4 01455 A4 01455 A4 01455 x2 x2 x2 ...
Page 7: ...Packing layout A4 02266 A4 01389 x2 A4 01591 A4 01391 A4 01415 ...
Page 8: ......
Page 12: ...1 1 2 Vacuum system ...
Page 37: ...Power supply mounting screws For better access to the screws ...
Page 50: ...Power supply screws For better access to the screws ...