Rear tanks hoses revision
Next step, in case the hoses are OK, is to check the filters above the tanks are not wet. These kind of
filters when get in contact with a liquid or if they have some type of particle block loose their
permeability.
If you find any dirt or debris inside of the filters,replace them.
The next step if the problem persists is to clean the suction duct of the applicators. For this purpose
use the cleaning syringe and a container full of warm water (cold also possible).
Cleaning syringe
Before carrying out the filling process, disconnect the tube with the red ring from the deposit tank as
shown in the image:
Suction tube disconnection
It is necessary for safety reasons, protecting the vacuum pump in case a treatment is started by
mistake.
To start the cleaning process, insert the blue syringe tube into the suction inlet of the hand piece.
Summary of Contents for Cooltech
Page 1: ...Cooltech SERVICE MANUAL Service Manual Cooltech V1 1 ...
Page 5: ... A4 01454 x2 A4 01453 A4 01454 A4 01455 A4 01455 A4 01455 x2 x2 x2 ...
Page 7: ...Packing layout A4 02266 A4 01389 x2 A4 01591 A4 01391 A4 01415 ...
Page 8: ......
Page 12: ...1 1 2 Vacuum system ...
Page 37: ...Power supply mounting screws For better access to the screws ...
Page 50: ...Power supply screws For better access to the screws ...