18. Troubleshooting
The Coag-Sense Meter continually checks its systems for
unexpected conditions. These may arise because of defective
components or consumables, environmental factors or due to User
handling and procedure errors. This section details how to resolve
most problems that you might encounter. If you have any
questions or problems during the troubleshooting process, note
the display wording and contact Technical Support at
866-903-
0890
or email
.
Meter
Display
Possible Cause(s)
Solution
ROOM TEMP
INCORRECT
SEE
MANUAL
The temperature of the room
is either below or above the
operating temperature range
of the meter.
Battery was depleted and
then plugged into AC adapter
while testing continued. The
charging of the battery can
generate enough heat to
raise the internal temperature
of the meter outside the
operating range.
Move the meter to a
place that is within the
operating temperature
range of the meter
(65ºF to 90ºF, 18ºC to
32ºC) and allow meter
time to adjust to correct
temperature.
Repeat
testing.
Allow
meter
to
complete charging prior
to initiating testing.
REMOVE
STRIP
Meter turned off with used
strip in it.
If no strip present, possible
shipment damage.
Remove the strip and
begin again.
Call Technical Support
WHEEL
PROBLEM
The test strip was not
inserted fully or may have
been inserted at an incorrect
angle or incorrect speed.
There may be a problem with
the wheel on the strip or with
the meter.
Reinsert
the
strip
holding the back of the
meter steady with one
hand while inserting the
strip completely using a
quick smooth motion
with the other hand. If
display
persists,
try
again with another new
strip.