Page 46
17. Cleaning and Disinfecting the Meter
No maintenance is required other than routine cleaning and/or
disinfecting.
When the power is off and the USB cable is not connected, the
meter housing can be cleaned and disinfected. Wipe all exposed
surfaces with Healthcare Bleach Germicidal Wipes containing
Sodium Hypochlorite (EPA No. 67619-12) for a contact time of 1
minute to pre-clean blood and other body fluids. Caution should be
taken to not get fluids inside the meter through the test strip port,
data transmission port or battery compartment. Dispose of the
used towelette. The meter should be allowed to air dry before use.
The test strip is designed to contain the patient sample, preventing
it from entering the meter. Do not clean/disinfect inside the meter
where the test strip is inserted. Cleaning this area should be
avoided. Please call Technical Support at 1-866-903-0890 if this
area requires cleaning/disinfecting.
WARNING
: Do not put the meter in liquid. Do not
allow liquids to get inside the meter or into any of
the connectors or plugs on the meter. If you
suspect any physical damage or deterioration of
the meter (such as cracking or gross distortion),
or if the meter does not turn on after cleaning,
call Technical Support.