24
S – Service, Calibration and Maintenance
Service and Unscheduled Repairs
The Pump must only be serviced or repaired by one of the following:
• The T34 distributor or its approved service representative,
• Qualified biomedical technicians trained and certified.
If the Pump is damaged in any way, return to authorized service personnel.
Unauthorized modifications or the use of any spare parts, other than those supplied by the
manufacturer or their distributor, will void any warranty.
Annual Maintenance and Calibration
Send the T34 Syringe Pump to the service center annually for maintenance and calibration.
Monthly Cleaning
Once a month (or as required) clean the lead screw thread (beneath the actuator) and guiding
rods with a small dry brush to remove debris or other particles.
Between Patients
Change the battery.
Before connecting the pump to a patient, and periodically during use, clean the unit using a lint-
free cloth lightly dampened with warm water and a mild detergent, disinfectant or 10% bleach
solution.
Always turn the syringe Pump off and remove the battery before cleaning.
Do not clean the Syringe Pump with chemicals such as Xylene, Acetone or similar solvents.
These chemicals can cause damage to components and labels. Use a lint-free cloth dampened
with warm water and mild detergent or 10% bleach solution.
Do not soak or immerse any part of the T34 in water or any other solution.
Storage
If the pump is to be stored for an extended period it should be cleaned and the battery removed.
Store in a clean, dry atmosphere at room temperature and, if available, use the original
packaging, or a suitable alternative, for protection.