22
Q – Specifications
Type:
Syringe Pump with motor driven linear actuator, pulsed motion
(540 pulses per mm)
Flow Rate:
0.05- 10ml/h in 0.01ml increments, 10 to 650ml/h in 1 ml increments
(1000 ml/hr available)
Actuator travel:
c.67 mm available
Syringe sizes:
2 ml to 50 ml (most commonly used manufacturers, see section T)
Accuracy:
±
5 % system accuracy (pump and set combined)
Occlusion pressure: 100-1500mmHg configurable (10mmHG increments).
Max. actuator force 50N (5 Kgf).
Battery:
9V alkaline, IEC 6LR61 type (or Lithium for extended life)
Battery Operation:
7 full deliveries depending on infusion & set up parameters.
Indicators:
4 Line LCD display (122 x 32 pixels) , Dual color operation LED
When a problem is detected, the T34 displays the following alarm
messages, sounds an audible alarm and the red LED lights: -
Alarms:
•
Occlusion or Syringe Empty
•
End Program
•
Near End
•
End Battery
•
Low Battery
•
Pump Paused Too Long
•
Syringe Displaced
•
Syringe Empty
T34 Dimensions:
169 x 53 x 23mm
Classification
Type CF Equipment (degree of protection against electrical shock)
Housing:
ABS (fire retardant)
Weight
210 gr. without battery
Electrical Safety
Complies with:
EN 60601-1
(Medical Electrical Equipment Safety),
IEC 60601-2-24
(Infusion pumps and controllers),
IEC 60601-1-4
(Programmable Electrical Medical System),
UL 60601- 1
and
CAN/CSA C22.2
No 601.1.
Standards
Manufacture in accordance to
ISO 9001
and
ISO 13485. CE marked
(In accordance with the Medical Devices Directive 93/42/EEC)
EMC
McKinley
T34 complies with IEC 60601-1-2 (EMC)
Environmental
Specifications
Non Operating Conditions
(Transportation and Storage): -
Temperature: -25
o
C to 55
o
C (-13
o
F to + 131
o
F)
Humidity: 5 % to 100% R.H., non-condensing
Air pressure: 48kPa to 110kPa
Operating Conditions:
The system may not meet all performance
specifications if operated outside of the following conditions: -
Temperature: +15
o
C to +45
o
C (+59
o
F to + 113
o
F)
Humidity: 20 % to 85% R.H. at +40
o
C, non-condensing
Air pressure: 70kPa to 110kPa