19
N - Change Set Up
An unlock code is required to enter the Change Set Up menu. Set up parameters should only
be changed by clinical or technical staff with healthcare provider code access rights.
When the pump is in program mode, or stopped during infusion, pressing the INFO
button.
Select Change Set Up from the menu using the UP / DOWN arrows and press YES
Pump will prompt for Healthcare provider Code (not given in this manual, please consult
senior clinical personnel or biomedical engineers), use the UP arrow to scroll code.
When the UP arrow is held down the pump will count up in single digits to ten, then in
tens to one hundred and then in hundreds thereafter. Scroll to the nearest point in tens
and then release the key. Press the key again, UP or DOWN arrow depending on digit
displayed on the screen, to select the final, correct code. Press YES to submit code.
Set up Menu
Parameter
Application
Range
Set Time & Date
Ensures all events in log are correctly date and
time stamped (ensure local update for DST)
FF Key Operation
Limits the forward movement of actuator caused
by pressing the FF key during syringe loading
0.1 – 100mm
Backlight Duration
Limits backlight duration to preserve battery life
0 – 60 secs.
Info Duration
Sets time info screens are displayed
1 – 20 secs.
Operation LED
Turns off the green operation LED to preserve
battery life. Red warning LED still operates
0 (Disabled) –
64 seconds
Titration Option
enables setting a titration limit (≤max.rate)
Enable/Disable
Default Duration
When entering a new program pump resets to this
default (setting to zero skips duration step during
programming). When Program Lock is ON (see
below) default duration cannot be set to zero.
00:00 (OFF) to
99:59 hours
Occlusion Pressure
Sets pressure level at which occlusion alarm will
activate
100–1500mmHg,
10mm increment
KVO Operation Rate
Activates Keep Vein Open infusion at End
Program
0 (KVO OFF) –
2.0ml/h
Program Lock
Prevents alteration of default duration and infusion
rate during set up (with Program Lock ON Default
Duration cannot be set to zero)
ON/OFF