background image

 

 

3

About blood pressure  

During each heart beat the arterial blood 
pressure varies between two utmost values: 
the systolic and the diastolic pressure. The 
peak pressure in the arteries is the systolic 
pressure and the lowest pressure is the 
diastolic pressure. In between these is the 
Mean Arterial Pressure (MAP) which is used 
to describe the average blood pressure.  

 

Oscillometric Method 

The blood pressure sensor can be used to determine blood pressure via the so-called 
oscillometric method. With this method, which is non-invasive, a cuff is placed around the 
arm and inflated by means of a pump, after which the cuff deflates via an automatic valve. 
The sensor measures the cuff pressure. This pressure varies because of the blood flow in 
the brachial artery.  
With inflation of the cuff, the external pressure on the artery rises, and hence the artery is 
increasingly compressed. At pressures exceeding the systolic blood pressure, the artery 
will be occluded. There are weak pressure pulses against the cuff in the rhythm of the heart 
beat which shows as small peaks in the graph. 
When the cuff is slowly deflated, the cuff pressure, and hence the external pressure on the 
artery will be lowered to that of the systolic blood pressure. Now, the artery is no longer 
continuously occluded. At systolic blood pressure, small amounts of blood pass through 
the compressed artery segment and cause changes in the artery volume, conducted to the 
cuff. This leads to pressure oscillations in the cuff. These oscillations increase with lower 
cuff pressure values, as more blood passes through the compressed artery. The maximum 
oscillation amplitude is reached around the mean arterial blood pressure. Then, as the 
pressure decreases until the cuff becomes fully deflated, the blood flow returns to normal 
and the oscillation amplitude decreases in the cuff decreases and small pulses remain at a 
low level below diastolic pressure.  
The point at which the largest oscillations are occurring corresponds to the mean arterial 
pressure (MAP). The point above the mean pressure at which the pressure difference grows 
rapidly correlates to the systolic pressure. The point below the mean pressure where the 
differences start to get small corresponds to the diastolic pressure. 
A similar method is used during the regular blood pressure measurement, a clinician, using 
the stethoscope, listens at the brachial artery for characteristics sounds of the pressure 
pulses (so-called Korotkoff sounds).  

 
Measurements with the Blood Pressure sensor

 

When performing blood pressure measurements, it is best to work with a partner. 
1.

 

Connect the Blood Pressure sensor to your interface. Attach the rubber hose from the 
cuff to the connector on the sensor.  

2.

 

Wrap the cuff firmly around your partner’s arm, approximately 2 cm above the elbow. 
The two rubber hoses from the cuff should be positioned over the bicep muscle 
(brachial artery) and not under the arm.  

Important: 

The person having his or her blood pressure measured must remain still 

during data collection—no movement of the arm or hand during measurements.

 

3.

 

Start your measurement in the Coach program.  

Figure 2.

 Pressure course of the heart beat in 

the brachial artery.  

Summary of Contents for 0377i

Page 1: ...BLOOD PRESSURE SENSOR 0377I 0 250 MM HG User s Guide CENTRE FOR MICROCOMPUTER APPLICATIONS http www cma science nl Figure 1 The Blood Pressure sensor...

Page 2: ...educational purposes Read this manual before you start measurements with the sensor Notice that over inflation of the cuff may cause pain and or injury Sensor specifications The Blood Pressure sensor...

Page 3: ...oscillations increase with lower cuff pressure values as more blood passes through the compressed artery The maximum oscillation amplitude is reached around the mean arterial blood pressure Then as th...

Page 4: ...ter clockwise The larger a subject s arm the slower the release valve will exhaust Helpful Tips Blood pressure readings will differ from person to person and even between measurements on the same indi...

Page 5: ...y 50 Diastolic blood pressure is calculated using the same method and the portion for p MAP usually 75 These percentage values are obtained from empirical research 2 The bell shaped envelope curve is...

Page 6: ...nd pressure by using the option Save as Determination of upper pressure values 1 Open the original measurement results 2 Repeat the steps 2 to 6 but this time mark the maximal pressure values the tops...

Page 7: ...p trend Name this quantity p difference Make this column also invisible 8 Create a new diagram with the name Envelope curve Select for the data range C1 Manual p trend Select for the data range C1 For...

Page 8: ...echnical data Pressure range 0 250 mm Hg Voltage output range 0 4 5 V Calibration function p mm Hg 64 94 Vout V Max pressure 1030 mm Hg without permanent damage Typical accuracy 3 mm Hg Temperature Co...

Reviews: