Description 

The Blood Pressure sensor is used to measure arterial blood pressure in humans (non-
invasively). It measures the pressure signal caused by the interaction between the cuff and 
the blood flow through the brachial artery.  
The pressure sensor uses the SenSym SDX05D4 pressure transducer. This element has a 
membrane that flexes as pressure changes. It is arranged to measure differential pressure. 
The sensor produces an output voltage that varies in a linear way with pressure. Special 
circuitry minimizes the errors that might be caused by changes in temperature. The sensor 
is delivered with standard adult size adjustable cuff (27 cm to 39 cm) and bulb pump (with 
release valve).  

 

The Blood Pressure sensor 

is equipped with a BT plug and can be connected to the 

following CMA interfaces: €Lab

CoachLab II/II

+

 and 

U

LAB. Furthermore the 

sensor can be used with Texas Instruments CBL

, CBL2

 and Vernier LabPro

.

 

 

 

 

Important:

 

The Blood Pressure sensor is not appropriate for medical applications.  

This sensor is to be used only for educational purposes. Read this manual before you start 
measurements with the sensor. Notice that over inflation of the cuff may cause pain and/or 
injury

.  

 

   

 

 

 
Sensor specifications  

The Blood Pressure sensor 

has a memory chip (EEPROM) with information about the 

sensor. Through a simple protocol (I

2

C) the sensor transfers its data: name, quantity, 

unit and calibration to the interface

1

 
Suggested experiments   

The following experiments can be performed with the Blood pressure sensor:  

 

Measure blood pressure before and after exercise and while sitting or standing; 

 

Compare blood pressure after voluntary isometric contractions (weight lifting) and a 
rhythmic activity such as running or biking; 

 

Study the effect of caffeine on blood pressure. 

 
Calibration 

The output of the Blood Pressure sensor is linear with respect to pressure. To collect data 
you can:  
1.

 

Use the calibration supplied by the sensor EEPROM memory.  

2.

 

Use the calibration supplied in the standard sensor library of the Coach program. The 
name of the Blood pressure sensor in the sensor library of Coach is  
Blood pressure sensor (0377i) (CMA) (0..250 mm Hg). 

 
 
 

                                                

 

 

1

 This is valid for the following interfaces: CMA €Lab

BT inputs of CoachLab II/II

+

 and 

U

LAB, 

TI CBL

and CBL2

, and Vernier LabPro.  

Summary of Contents for 0377i

Page 1: ...BLOOD PRESSURE SENSOR 0377I 0 250 MM HG User s Guide CENTRE FOR MICROCOMPUTER APPLICATIONS http www cma science nl Figure 1 The Blood Pressure sensor...

Page 2: ...educational purposes Read this manual before you start measurements with the sensor Notice that over inflation of the cuff may cause pain and or injury Sensor specifications The Blood Pressure sensor...

Page 3: ...oscillations increase with lower cuff pressure values as more blood passes through the compressed artery The maximum oscillation amplitude is reached around the mean arterial blood pressure Then as th...

Page 4: ...ter clockwise The larger a subject s arm the slower the release valve will exhaust Helpful Tips Blood pressure readings will differ from person to person and even between measurements on the same indi...

Page 5: ...y 50 Diastolic blood pressure is calculated using the same method and the portion for p MAP usually 75 These percentage values are obtained from empirical research 2 The bell shaped envelope curve is...

Page 6: ...nd pressure by using the option Save as Determination of upper pressure values 1 Open the original measurement results 2 Repeat the steps 2 to 6 but this time mark the maximal pressure values the tops...

Page 7: ...p trend Name this quantity p difference Make this column also invisible 8 Create a new diagram with the name Envelope curve Select for the data range C1 Manual p trend Select for the data range C1 For...

Page 8: ...echnical data Pressure range 0 250 mm Hg Voltage output range 0 4 5 V Calibration function p mm Hg 64 94 Vout V Max pressure 1030 mm Hg without permanent damage Typical accuracy 3 mm Hg Temperature Co...