background image

 

 

5

 

The blood pressure increases with the age. The rule of thumb for the normal systolic 
pressure is the formula 100 + age. 

 

Blood pressure values are rounded to 5 mmHg. 

 
Calculating the blood pressure in the Coach software 

The blood pressure cannot be read directly from the measurement graph. A detailed 
description of the data processing is given in this chapter. 

 

The result of the measurement with the Blood pressure sensor is a pressure - time graph in 
which the pressure pulses of the blood are superposed on the decaying trend pressure of the 
cuff. The pressure pulses, when extracted from the cuff pressure, form an oscillating 
waveform. The peak-to-peak amplitudes of this waveform create a bell shaped “envelope 
curve” vs. pressure. Within the envelope, the pulse amplitudes increase through the 
systolic blood pressure and continue increasing until a maximum amplitude (Mean Arterial 
Pressure - MAP), is reached. After that the pulse amplitudes decrease further until hardly 
noticeable.  
Generally, the systolic blood pressure is calculated by determining the point along this 
bell-shaped envelope for which p>MAP using a known percentage of the maximum 
amplitude (usually 50%). Diastolic blood pressure is calculated using the same method and 
the portion for p<MAP (usually 75%). These percentage values are obtained from 
empirical research.

2

 

The bell shaped “envelope curve” is extracted from the measurement graph as the 
difference graph of the pulse pressure graph (the maximal values of the pressure pulses) 
and the trend pressure graph. 

 

Important:  

Save your original blood pressure 
measurements. These are needed for the 
data processing.  

Determination of the cuff pressure 

1.

 

Determine the number of the 
measured points in your 
measurement (from the 
Measurement settings or from the 
Measurement table). 

2.

 

Select in the diagram window menu 
the 

Process > Select/Remove data 

option

.

 

3.

 

Select the method 

Point-by-point

4.

 

Mark the very first point, very last point and points where the pressure values are 
locally minimal. These points indicate the graph of the deflating pressure trend of the 
cuff (i.e. without the pulses). It is enough to take every second minimum (see figure 4). 

Tip: 

Use the cursor key to go through points. Find the point where the pressure value 

                                                

 

 

2

 

Mehlsen J et al (1999). Vejledende retnigslinier for hjemmeblodtryksmåling. Ugesk Læ Klaringsrapport 

nr. 8, 1999. Online beschikbaar op http://www.hypertension.suite.dk/klar899.pdf

  

Figure 4. 

Select the points at which the pressure value 

is minimal, just before the pressure signal rises again.

Summary of Contents for 0377i

Page 1: ...BLOOD PRESSURE SENSOR 0377I 0 250 MM HG User s Guide CENTRE FOR MICROCOMPUTER APPLICATIONS http www cma science nl Figure 1 The Blood Pressure sensor...

Page 2: ...educational purposes Read this manual before you start measurements with the sensor Notice that over inflation of the cuff may cause pain and or injury Sensor specifications The Blood Pressure sensor...

Page 3: ...oscillations increase with lower cuff pressure values as more blood passes through the compressed artery The maximum oscillation amplitude is reached around the mean arterial blood pressure Then as th...

Page 4: ...ter clockwise The larger a subject s arm the slower the release valve will exhaust Helpful Tips Blood pressure readings will differ from person to person and even between measurements on the same indi...

Page 5: ...y 50 Diastolic blood pressure is calculated using the same method and the portion for p MAP usually 75 These percentage values are obtained from empirical research 2 The bell shaped envelope curve is...

Page 6: ...nd pressure by using the option Save as Determination of upper pressure values 1 Open the original measurement results 2 Repeat the steps 2 to 6 but this time mark the maximal pressure values the tops...

Page 7: ...p trend Name this quantity p difference Make this column also invisible 8 Create a new diagram with the name Envelope curve Select for the data range C1 Manual p trend Select for the data range C1 For...

Page 8: ...echnical data Pressure range 0 250 mm Hg Voltage output range 0 4 5 V Calibration function p mm Hg 64 94 Vout V Max pressure 1030 mm Hg without permanent damage Typical accuracy 3 mm Hg Temperature Co...

Reviews: