Sometimes it is necessary to re-calibrate the 4
values. This can be easily achieved by following the below procedure.

1.

 

Apply a fixed accurate 4mA source to the current loop interface.

2.

 

Dropdown the 'Calibrate Loop' menu and select 'Set 4mA Point'.

3.

 

Change the source current on the interface to an accurate 20mA.

4.

 

Dropdown the 'Calibrate Loop' menu and select the 'Set 20mA Point'.

5.

 

Change the loop current to any random value between 4mA and 20mA.

6.

 

Press the 'Get Current Setup' button and read back the settings.

7.

 

This random loop current should now display correctly in the 'General Details' box as 'Loop Current'.

8.

 

Disconnect loop current. 

Upgrading firmware on the logger can be done through the Bluetooth interface. Newer versions of the software for each 
type of device is embedded into the Cloudworks software so will not need an internet connection to collect the data.

When in Command Mode, press the 'Upgrade Firmware' button. The first thing the software  will do is to determine  the 

current firmware version on the logger and see if 
be prompted as such and given an option to re
you would like to try again. 
Before continuing, the following prompt will flash on the screen.
 

Please do not at any stage interrupt the process while the firmware is downloading as this may leave the logger inoperable 

and necessitate that it get sent back to the manufacturer for recovery.

 

 

 
Hardware Test 

 
 
 
Setup Bluetooth 

 
 
 

 
 
 
 

Calibrate Loop 
 
Upgrade Firmware 
 

 

 
Dropdown menu to activate various hardware tests. Selecting these options 

will send the appropriate command to the logger and display the result in the 
diagnostic screen.
 
Bluetoot

current Device Code for the logger. This makes it easier to select the correct 
device when connecting to a Bluetooth interface. Should the Device Code not 
appear  on  the  list  of  devices  when 

(legacy devices), then pressing this button will reset the interface and correct 
the  naming  system.  This  function  can  also  be  used  to  reset  the  Bluetooth 
interface to the correct setup.
 

Used to calibrate the 4
 
The onboard firmware can be upgraded through the Bluetooth interface. The 
process is explained in further detail below.

MANUAL : CDS541 CUMULUS PULSE LOGGER 

NAME 

CLASSIFICATION 

DATE 

PAGE

calibrate the 4-20mA loop to correct it or to re-calibrate for a completely different set of 

values. This can be easily achieved by following the below procedure. 

Apply a fixed accurate 4mA source to the current loop interface. 
Dropdown the 'Calibrate Loop' menu and select 'Set 4mA Point'. 

nt on the interface to an accurate 20mA. 

Dropdown the 'Calibrate Loop' menu and select the 'Set 20mA Point'. 

Change the loop current to any random value between 4mA and 20mA. 
Press the 'Get Current Setup' button and read back the settings. 

current should now display correctly in the 'General Details' box as 'Loop Current'.

Upgrading firmware on the logger can be done through the Bluetooth interface. Newer versions of the software for each 

into the Cloudworks software so will not need an internet connection to collect the data.

When in Command Mode, press the 'Upgrade Firmware' button. The first thing the software will do is to determine the 

current firmware version on the logger and see if there is an upgrade available. Should the firmware be up to date, you will 
be prompted as such and given an option to re-flash the firmware. Re-flashing can happen if your last upgrade failed and 

g prompt will flash on the screen. 

 

 

Please do not at any stage interrupt the process while the firmware is downloading as this may leave the logger inoperable 

and necessitate that it get sent back to the manufacturer for recovery. 

Dropdown menu to activate various hardware tests. Selecting these options 

will send the appropriate command to the logger and display the result in the 
diagnostic screen. 

Bluetooth  interfaces  are  set  to  default  'Friendly  Names'  that  include  the 

current Device Code for the logger. This makes it easier to select the correct 
device when connecting to a Bluetooth interface. Should the Device Code not 
appear  on  the  list  of  devices  when  connecting  to  the  Bluetooth  interface 

(legacy devices), then pressing this button will reset the interface and correct 
the  naming  system.  This  function  can  also  be  used  to  reset  the  Bluetooth 
interface to the correct setup. 

Used to calibrate the 4-20mA. This is explained in further detail below.

The onboard firmware can be upgraded through the Bluetooth interface. The 
process is explained in further detail below. 

 

PAGE 

for a completely different set of 

current should now display correctly in the 'General Details' box as 'Loop Current'. 

Upgrading firmware on the logger can be done through the Bluetooth interface. Newer versions of the software for each 

into the Cloudworks software so will not need an internet connection to collect the data.

 

When in Command Mode, press the 'Upgrade Firmware' button. The first thing the software  will do is to determine the 

there is an upgrade available. Should the firmware be up to date, you will 

flashing can happen if your last upgrade failed and 

Please do not at any stage interrupt the process while the firmware is downloading as this may leave the logger inoperable 

Dropdown menu to activate various hardware tests. Selecting these options 

will send the appropriate command to the logger and display the result in the 

h  interfaces  are  set  to  default  'Friendly  Names'  that  include  the 

current Device Code for the logger. This makes it easier to select the correct 
device when connecting to a Bluetooth interface. Should the Device Code not 

connecting  to  the  Bluetooth  interface 

(legacy devices), then pressing this button will reset the interface and correct 
the  naming  system.  This  function  can  also  be  used  to  reset  the  Bluetooth 

This is explained in further detail below. 

The onboard firmware can be upgraded through the Bluetooth interface. The 

Summary of Contents for CDS541

Page 1: ...se Logger User Manual User 28 February 2018 V2 0 Draft Confidential CDS541 Cumulus Pulse Logger User Manual docx This document contains information relating to the rights of Cape Digital Solutions The...

Page 2: ...021 MANUAL CDS541 CUMULUS PULSE LOGGER NAME CLASSIFICATION DATE PAGE AMENDMENT HISTORY Amendment Details Amended By Compile document Alex Romanov Updated modem technology specs Alex Romanov New config...

Page 3: ...0mA loop input and 24V loop power generator 1 8 5 Using the Pulse Input with echo output 1 8 6 Using the ECO interface 2 Logger Setup 2 1 Setup via Bluetooth Configure 2 2 Setup via Remote Configure 2...

Page 4: ...r easy installation and setup without the need for a complicated server infrastructure Setup and management software CloudWorks is supplied free of charge to allow clients to manage their own networks...

Page 5: ...of 30mA allowing for the powering of an external piece of equipment or current loop device eliminating the need option will also function under battery operation 2 x Pulse inputs with echo output Puls...

Page 6: ...NAME CLASSIFICATION DATE PAGE or Bluetooth Interface Central Processor GPS Satellite Receiver Input Output Interface 3 x Alkaline Batteries External 12V Power Supply Power Management 2 wire Isolated 4...

Page 7: ...usly connected to the server allowing for the streaming Cumulus Pulse logger uses standard D Cell alkaline batteries that will last several years In this mode the logger will wakeup adjustable normall...

Page 8: ...und 13 Pulse 2 input 14 Ground for Pulse 2 input 15 Ground for Pulse 2 output pulse splitter output 16 Pulse 2 pulse splitter output switch to ground 17 Digital input 18 Ground for digital input 19 EC...

Page 9: ...also populating the logger with alkaline batteries means that the logger can be put e is solar power available the logger will remain online with the server Should the solar power fail the logger wil...

Page 10: ...igned to connect to a volt free mechanical or electronic switch The logger has Example for connecting to a volt free contact Example for connecting to an Open Collector electronic switch note that pin...

Page 11: ...has been designed to supply a maximum of 30mA Please note using the 24V power generator to power a loop device uses a substantial amo he loop should only be powered each time the device needs to be re...

Page 12: ...nfiguration of MANUAL CDS541 CUMULUS PULSE LOGGER NAME CLASSIFICATION DATE PAGE Example configuration of water meter with electronic interface Please note the polarity of the interface Ground connecti...

Page 13: ...umulus Pulse logger has an onboard Bluetooth interface which allows for the device There are two different places in the software to activate the Bluetooth connection The first option is from the log...

Page 14: ...r the Refresh button to start the search again Please remember that battery operated devices will need the Bluetooth interface activated before it will be found by the Bluetooth Server This is normall...

Page 15: ...Firstly it will stream diagnostic information is operating This will be information like the GSM signal strength next wakeup time is initiating a connection to the server ommand Mode which allows the...

Page 16: ...ed into Command Mode and the current setup data retrieved and displayed on the right hand screen nt to the device only after the parameter setup field has been left What this means is that while adjus...

Page 17: ...an be cleared by right clicking in this box and before any of the settings changes or other instructions are accepted command in Diagnostic Mode will cause the logger to switch to Command Mode but tha...

Page 18: ...attery or Mains Digital input status 20mA loop current reading in milliamps 20mA current loop reading with its calibration data implemented including the unit The I O Parameter block is the data used...

Page 19: ...pulate the database that is associated with a specific logger Each database and Vendor Code assigned to it Data is matched between these systems using that code This code is also used to separate comm...

Page 20: ...number is normally not required as your network will default this value This can be customised using the SMS Serv Centre field If you are using a foreign simcard on a network the logger will need to b...

Page 21: ...gle chance to wake up fails it will return to sleep and a only reconnect at the next wakeup period At the start of a wake up the Max Start Time timer will be started Should the connection to the serve...

Page 22: ...V 30mA power generator The Cumulus Pulse Logger can generate a stable 24V allowing for the powering of an external piece of equipment or current loop device the need for a secondary power supply This...

Page 23: ...for a couple of minutes typically 5 minutes set in the server software in order to get a fix If this takes too long the remote device will be put back to sleep without updating the position This is do...

Page 24: ...ent Upgrading firmware on the logger can be done through the Bluetooth interface Newer versions of the software for each into the Cloudworks software so will not need an internet connection to collect...

Page 25: ...is complete the following message will be displayed Please power cycle the logger and then reconnect the Bluetooth interface if you wish to carry out further configurations when a logger is in battery...

Page 26: ...when you configure via the Bluetooth setup option Again d commands are sent to the device only after the parameter setup field has been left What this means is that while adjusting a parameter in its...

Page 27: ...enting you from making any further changes The Remote Configure screen has one further tool available in its arsenal and that is the Send Test SMS function The purpose of this function is to not only...

Page 28: ...n which are picked up As you can see the parameters are the same as when you configure via the Bluetooth setup option with one or two minor All the parameter fields are empty at this point To create a...

Page 29: ...ts that are still present in the database Each command has associated with it a status those that still need to be executed Listed also is the response to each command for diagnostic purposes You can...

Reviews: