Models DC-1, -10, -20B DC Spark Testers
DC-1, -10, -20B Instruction Manual
–
Page 48
18.
Press the “R” button. The Fault Count display should show “
S
.
00
.
0
”
representing the set point value.
19.
Using the UP/DOWN arrows adjust the output voltage up so that the Fault
Count display shows “
S
.
01
.
0
” for 1kV.
20.
Press the “R” button again to return the Fault Count Display to display the
Voltage Monitor Value. If this Value does not match the output voltage value on
the EVM
, press the “CR” button to change the fault Display to “
*
.
01
.
0
”.
21.
Using the UP/DOWN arrows adjust the reading on the display to match the
match the output voltage value on the EVM.
22.
Once the Voltage monitor value matches the output voltage value on the EVM,
press the “R” button to return the Fault Count Display to read the set point
value “
*
.
01
.
0
”.
23.
Repeat these steps until the all calibration values are within tolerance.
24.
Once calibration is complete press the “A” button once to exit the Manual
calibration mode.
Maintenance
Fuses
The fuses in this equipment are not expected to fail in normal operation. Their
failure may be an indication of equipment malfunction requiring qualified repair
personnel.
There is one fuse associated with the spark tester’s operating voltage, located in
the ON/OFF power switch on the back panel of the unit.
Two additional fuses that could be defective are found on the relay output
printed circuit board, behind the back panel.
Fuse
Relay Fuse