Models DC-1, -10, -20B DC Spark Testers
DC-1, -10, -20B Instruction Manual
–
Page 30
Output Functions are designed to relay information to an alarm, PLC, etc. The
Output functions can be assigned to output pins labeled
‘OUT 0, OUT 1, OUT 2,
and OUT 3.
General Outputs
Output Disabled (OFF)
This output will not do anything, under any condition.
Cover Open (CVO):
This output will change state when the safety interlock cover is open. If the
safety interlock cover is closed, the spark tester is able to produce high voltage.
System OK (SOK)
This output will change state when there is a hardware malfunction that could
cause the unit to operate outside of acceptable limits.
Discharge Bar (DIS)
This output is the discharge bar output. When enabled, it will be activated
when the voltage is enabled and the voltage setting is above 0 and disabled
when the voltage is set to 0 or the voltage is disabled.
High Voltage Enable (HVE)
This output will change state when the unit meets all conditions to generate
high voltage.
Voltage Monitor Output Options
High Voltage On Lamp (HVL)
This output will change state when the actual voltage is equal to or above the
set
“HVON”
configuration item value, and will turn off when the voltage is equal
to or below the set
“HVOF”
configuration item value.
–
High Voltage On
Set in kV. If the actual voltage goes above this set point, any
“HVL” output will
change state.
–
High Voltage Off
Set in kV. If the actual voltage drops below this set point, any
“HVL”
output will change state.