3. Number of separators
The measurement of air quantity is based on meas-
uring the chamber pressure. To achieve the correct
result, the number of separators must be defined for
the logic. The hood can have 1–6 separators when it is
a one-sided ‘wall model’.If it is two-sided, there can be
1–12 separators.
Define the number of separators in the configuration
to the logic. If there are ‘blinds’ in the hood, they are
not included in the number of separators. For exam-
ple, the maximum number of separators in a hood
with a width of two metres is 4.
4. Approve the selections by clicking
“Commission selections”, after which the device will
restart. After this, the hood will be ready to use. The
UV lamp switches on automatically when the cham-
ber’s pressure is over 10 Pa and the SlideSafe switch is
in the OFF position. When the UV lamp is on, the lamp
symbol is shown in the screen’s top row.
For other settings than the hood controller’s basic settings, please read through the separate user manual,
downloadable from
climecon.fi/clm
.
The Climecon CleanMaster system can also be connected to the external ModBus control device.
In this case, a Slave address in the range of 1–250 will be defined for all devices. The ModBus parameter
map is available through
Climecon
Oy’s customer service.