background image

10 

 

Screen 
display 

Operating Instructions 

Displayed State 

Sound and backlight 

Switching between ºF/ºC 

 

 

 

In power-off state, press and hold 
the 

Ear 

button for 6-7 seconds to 

enter the temperature switching 
mode. "

ºC

" or "

ºF

" are flashing. 

Press the 

Ear 

button to switch 

between ºC and ºF.The 
thermometer shut down after 6 
seconds, set up successfully. 

silence 

Error information & low battery

 

 

 
The ambient temperature is higher 
than 40.0ºC (104.0ºF) or lower than 
10.0ºC (50.0ºF). 

3 short beeps and 
yellow backlight for 3 
seconds. 

 

An error occurs when data is being 
read from or written into the 
memory, or the temperature 
correction is not complete. Please 
contact your supplier. 

3 short beeps and 
yellow backlight for 3 
seconds. 

 

When the battery voltage is lower 
than 2.51V ± 0.15V, the low battery 
symbol will appear on the display 
after turning the device on. Please 
replace the batteries. 

silence 

Power Off Mode 

In any mode, if no operation for 10 seconds, the thermometer will power 
off automatically. 

 
Attention: 

   

You must wait for the backlight to turn off before measuring 

again.

 

    Electromagnetic interference: contains sensitive electronic 

11 

 

components and should not be used in the presence of large 
electromagnetic interference, (such as near mobile phones and 
microwaves) 

    Please dispose of the product and batteries in accordance with 

local regulation requirements. 

 
 

8. Cleaning and Disinfection 

Cleaning 

1. 

Clean the temperature probe with 
a soft cloth. Clean the lens of the 
temperature probe with a cotton 
swab. 

 
2. 

Wipe the thermometer body with a 
slightly damp soft cloth, and gently 
dry the body with a piece of tissue 
paper. 

 

    Keep water off the lens during the cleaning process. Otherwise, 

the lens may be damaged. 

    The lens may be scratched if it is cleaned with a piece of tissue 

paper, resulting in inaccurate readings. 

    Do not clean the thermometer with corrosive cleansers. During 

the cleaning process, do not touch the lens using hard objects, 
immerse any part of the thermometer into liquid, or allow liquid to 
enter the thermometer. 

Disinfection 

1. 

Disinfect the thermometer body and the area around the 
temperature probe with a cloth slightly moistened with 75% 
medical alcohol. 

    Do not use hot steam or ultraviolet radiation for disinfection. 

Otherwise, the thermometer may be damaged or age quickly. 

 

9. Maintenance 

1. 

After each use, clean the temperature probe as described in 

Summary of Contents for CLETEF-786

Page 1: ...CLETEF 786 Infrared Thermometer Instruction Manual 0482...

Page 2: ...n be used to read body temperature by measuring the ear or forehead temperature and is suitable for both adult and children NOTE temperature of children below 3 months cannot be accurately measured wi...

Page 3: ...cklight Information 7 7 Display and Operation Instructions 8 8 Cleaning and Disinfection 11 9 Maintenance 11 10 Replacing the Batteries 12 11 Troubleshooting 12 12 Technical Specifications 14 13 After...

Page 4: ...Simple Diagnostics Inc Address P O Box 128 Williston Park NY 11596 USA Tel USA 877 342 2385 Tel Inter 1 954 603 0572 Website www SimpleDiagnostics com Email Sales SimpleDiagnostics com...

Page 5: ...ode design you are free to take body temperature by detecting the infrared heat given off by the forehead or eardrum The forehead temperature measurement range is from 89 6 to 108 0 32 to 42 2 The ear...

Page 6: ...ration distance in meters m b Field strengths from fixed RF transmitters as determined by an Electromagnetic site survey a should be less than the compliance level in each frequency range b Interferen...

Page 7: ...ipment should be used no closer to any part of the CLETEF 786 Infrared Thermometer including cables than the recommended separation distance calculated from the equation applicable to the frequency of...

Page 8: ...s test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The CLETEF 786 Infrared Thermometer uses RF energy only for its internal function Therefore its RF emissions are ve...

Page 9: ...n provided for in the 5 4 How to Use Forehead Mode Rapid measurement 1 As shown in Figure A put the cover on the probe and position the thermometer at the center of the forehead just above the eyebrow...

Page 10: ...used within 10 1 seconds Outer dimensions mm 146 5 40 52 4 Weight g Thermometer with batteries 85 0 g Liquid Ingress Protection IPX0 The date of production See the label Life 5 years The infrared the...

Page 11: ...ion No Range Sounds Backligh t Meaning Ear temperature mode 1 0 37 4 32 0 99 4 a long beep Green Acceptable 2 37 5 37 9 99 5 100 3 a long beep Yellow Elevated 3 38 0 100 0 100 4 212 0 6 short double a...

Page 12: ...xceeds 108 0 F 42 2 C in forehead temperature mode 3 short beeps with red backlight for 3 seconds The measured value falls below 32 0 F 0 C in ear temperature mode The measured value falls below 89 6...

Page 13: ...er is faulty If the warranty period has not expired contact us Only the battery symbol is displayed The battery level is low Replace the Batteries AAAx2 Er1 is displayed The ambient temperature is out...

Page 14: ...erference contains sensitive electronic 11 components and should not be used in the presence of large electromagnetic interference such as near mobile phones and microwaves Please dispose of the produ...

Reviews: