3. Product Design
(1) Probe cover (put the cover on when using forehead mode)
(2) Probe (take off the cover when using ear mode)
(3) Head button
(4) LCD display screen
(5) Ear button
(6) Battery cover
Display description
1. Ambient temperature
2. Ready for measurement
3. Ear mode
4. Forehead mode
5. Temperature unit (
℃
/
℉
)
6. Low power indicator
7. Memory mode
8. Temperature value
17
ACCOMPANYING DOCUMENTS.
● Portable and mobile RF communications equipment can affect
CLETEF-786 Infrared Thermometer
● The CLETEF-786 Infrared Thermometer should not be used
adjacent to or stacked with other equipment.
Guidance and manufacturer´s declaration – electromagnetic
emission –for all EQUIPMENT AND SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic emission
The CLETEF-786 Infrared Thermometer is intended for use in the
electromagnetic environment specified below. The customer or the user of
the CLETEF-786 Infrared Thermometer should assure that it is used in such
an environment.
Emissions
test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The CLETEF-786 Infrared Thermometer
uses RF energy only for its internal
function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The CLETEF-786 Infrared Thermometer
is suitable for use in all establishments,
including domestic establishments and
those directly connected to the public low-
voltage power supply network that
supplies buildings used for domestic
purposes.
Guidance and manufacturer's declaration – electromagnetic
immunity –for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The CLETEF-786 Infrared Thermometer is intended for use in the
electromagnetic environment specified below. The customer or the user of
the CLETEF-786 Infrared Thermometer should assure that it is used in such
an environment.
Immunity test
IEC
60601
test level
Compliance
level
Electromagnetic
environment- guidance
Electrostatic
discharge
(ESD)
±6 kV
contact
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with