background image

15/11/2018 

 

Page 9 

Tape 

1. Autoclave or plastic backed general tape should be used.  A length 

5cm longer than the width of each end of the tray should be cut. One 
length should be placed over one end of the tray and stuck m1cm in 
from the tray edge. This should then be folded, and the edges sealed 
securely. Repeat for the other end and place onto a level surface for 
gel pouring.  

2. Place the comb(s) in the grooves. Each tray has more than one comb 

grove so that multiple combs can be used. Using multiple combs 
increases sample number available per gel but decreases run length 
and care must be taken to ensure that samples from the first wells do 
not migrate into the lanes of the second comb wells. 

3. Pour in the agarose carefully so as not to generate bubbles. Any 

bubbles that do occur can be smoothed to the edge of the gel and 
dispersed using a pipette tip. 

4. Allow the agarose to set, ensuring that the gel remains undisturbed. 

5. Carefully remove the gel casting gates and comb and transfer the gel 

including tray to the main tank. 

Running the Gel 

1. Mix the sample to be loaded with sample buffer – see solutions for 

common sample buffers. Usually 3ul of sample buffer is adequate but 
less may be used with sample volumes of less than 10ul. 

2. Fill the unit with buffer until the gel is just flooded with buffer. This will give 

the fastest resolution times. For enhanced quality of resolution of 
sample, fill the unit to 5mm above the gel.  

3. Load the samples into the wells using pipettes. Multi-channel pipettes 

can be used for loading samples with MC compatible combs, see 
listing in accessories for identification of these. 

4. Carefully place the lid on the tank and connect to a power supply. 

5. Typically, gels are run at between 90 and 150 volts. However, maximum 

voltages are indicated on the serial badge of each unit. It should be 
noted that higher voltages generally give faster but poorer quality 
sample resolution. 

Summary of Contents for MSSCREEN-TRIO

Page 1: ...multiSUB Screen Horizontal System...

Page 2: ......

Page 3: ...RIO MSSCREEN16 NC MSSCREEN24 NC MSSCREEN32 NC MSSCREEN TRIO NC Record the following for your records Model _____________________ Catalogue No _____________________ Date of Delivery ___________________...

Page 4: ...ons 6 Setting up the Horizontal Gel Tanks 6 Fitting Electrodes 6 Fitting Loading Guides 6 Gel Preparation 7 Gel Pouring 8 Flexicaster 8 Tape 9 Running the Gel 9 Gel Staining and Viewing 10 Solutions 1...

Page 5: ...he complete manual thoroughly The unit must never be used without the safety lid correctly in position The unit should not be used if there is any sign of damage to the external tank or lid These unit...

Page 6: ...x 16cm W x L 6 x MS26 28MC 1 1mm 28 sample MS26 LG Strips MS26 WP Platform CSL CAB None MSSCREEN24 NC MS26 UV24 26 x 24cm W x L 6 x MS26 28MC 1 1mm 28 sample MS26 LG Strips MS26 WP Platform CSL CAB No...

Page 7: ...ions Power supply cables 1500V rated with retractable 4mm connectors Safety lid and viewing pane High impact acrylic construction Height adjustable combs Acrylic UV transparent gel tray UV transparent...

Page 8: ...Fitting Electrodes Note the position of the lid on the unit This shows the correct polarity and the correct orientation of the cables black is negative and red positive Remove the lid from the unit N...

Page 9: ...ose powder to a conical flask 2 Add the appropriate amount of 1x TAE or TBE solution from the table above To prevent evaporation during the dissolving steps below the conical flask should be covered w...

Page 10: ...end of the Flexicaster 3 Position the movable end of the Flexicaster so that the silicone mat is pushed against the other end of the tray 4 Turn the cam so that the silicone mat tightly seals against...

Page 11: ...ip 4 Allow the agarose to set ensuring that the gel remains undisturbed 5 Carefully remove the gel casting gates and comb and transfer the gel including tray to the main tank Running the Gel 1 Mix the...

Page 12: ...sed 3 Rinse the gel twice for a couple of seconds with distilled water 4 Transfer the gel to a UV Transilluminator The samples will often appear as brighter clearer bands when photographed or viewed u...

Page 13: ...lycerol 0 25 bromophenol blue and 0 25 xylene cyanole FF in 1x TAE buffer Only 1 10 ml of the 10x loading dye should be prepared Ethidium Bromide Solution Add 10 mg of Ethidium Bromide to 1 ml distill...

Page 14: ...running surface Curved line or distortion of bands Bubbles in sample wells Remove bubbles prior to electrophoresis Curved bands smiles Sample overload Reduce load Differential relative mobilities Sam...

Page 15: ...gents include dishwashing liquid Hexane and Aliphatic hydrocarbons The units should not be left to in detergents for more than 30 minutes The units should never come into contact with the following cl...

Page 16: ...C 0 75mm thick MS26 56MC 0 75 Comb 56 sample MC 0 75mm thick MS26 28MC 1 Comb 28 sample MC 1mm thick MS26 56MC 1 Comb 56 sample MC 1mm thick MS26 28MC 1 5 Comb 28 sample MC 1 5mm thick MS26 56MC 1 5 C...

Page 17: ...5g tablets CSL RUNSAFE CSL RUNSAFE Package 1 ml vial CSL TBEP Powdered Tris Borate EDTA Running Buffer 10 sachets 1litre pk TBE10X1L Cleaver Buffer Tris Borate EDTA Running Buffer 10 x 1L TBE10X5 Cle...

Page 18: ...n performed by anyone other than CSL or an appointed distributor or representative are no longer under warranty from the time the unit was modified Units which have accessories or repaired parts not s...

Page 19: ......

Page 20: ......

Reviews: