process is indicated by the red LED.
4. Having the battery charged first of all disconnect
the battery and then disconnect the charger from
the mains.
5. The battery has to be charged in temperatures
between 0 – 45°C
Do not leave the battery charger connected to the
mains when not used.
6.2.2. Battery replacement
CA Chemical 3F
see Picture appendix no.4
Picture 4A :
Taking the replaceable battery off the powered respi
-
ratory unit : grasp the unit from the front so that you
rest your index fingers against the safety locks of the
battery to release them. Push your thumbs into the rims
at the back of the sides of the battery. Then by pulling
both thumbs away from the unit you will release the
battery and push it off the guide ways.
Picture 4B :
Pushing the battery onto the powered respiratory unit:
push the battery into the guide ways on the bottom side
of the unit.
Push the battery towards the unit until both safety locks
click into the right position and safely lock the battery.
6.2.3. Spare parts and accessories
The manual for assembling spare parts and acces-so-
ries is presented in the Picture appendix in the chapter
called Accessories.
7. Possible faults
In case of any fault, a sudden decrease or increase of
air supply when the user is in a contaminated working
area, it is vital to leave the place and to check the fo
-
llowing:
• Whether the unit is correctly assembled.
• Battery charging.
• The function of the battery charger.
• Filter clogging.
• Whether the air hose is not damaged. It is necessary
to make sure that the air hose does not get caught
on protruding objects and that there will not originate
a rupture during work.
• Whether the sealing of the safety hood is all right.
• In case of working time reduction at one charging it
is necessary to replace the battery.
8. Storage
All parts of the CleanAIR
®
system must be stored at
temperatures between -10°C and 55°C with relative air
humidity inbetween 20 and 95 % Rh. Maximum storage
life in undamaged wrapping is 2 years excluding batte-
ries! During storage self-discharge of batteries occurs.
Batteries stored for long terms must be fully charged
every 6 months.
9. Warranty
The Quality Warranty ensures you will receive a repla-
cement if a product presents any manufacturing or
material defects within 12 months from the date of
purchase. For batteries it is 6 months from the date of
purchase. The warranty claim must be lodged at sales
organization. At the same time the proof of purchase
must be handed in ( invoice or certificate of delivery).
Fault
Probable cause
Recommendation
The unit does not work at all
Battery discharge
(check : the unit starts working
after replacing the battery with
another functional battery)
Fault of motor, electronics or
charging connector
Recharge the battery.
(if the problem outlasts, check the
battery)
Return to the producer for repairing.
The unit does not supply satisfactory
amount of air.(low air-flow)
Blocked air-hose or air channel.
Air leakage.
The battery is not charged
enough.
Clogged filter.
Check and remove the possible block.
Check all sealing elements and joints, make
sure the hose is not damaged and without
leakage.
Charge the battery.
(if the problem outlasts, check the battery)
Replace the filter.
The unit works only for a short time.
Clogged filter.
The battery is not charged
enough .
Replace the filter.
Charge the battery.
(if the problem outlasts, check the battery)
It is impossible to charge the battery..
Fault of charger.
Bad battery connector .
Contact your supplier.
Check the battery contact.
It is impossible to charge the battery
sufficiently.
Termination of battery life time.
Install a new charged battery.
The warranty can be acknowledged only if no inter-
-ventions into the powered respiratory unit and char-
-ger have been made. The warranty does not cover
defects caused by the late replacement of the filter or
by using a filter which was damaged by cleaning or by
filter blow-through.
CA CHEMICAL 3F
3F Plus
Minimum air-flow
Display
Weight of filter unit excluding filters
Filter unit noise
Battery life time
One recharging cycle
Belt size
Advised temperature range for work
Advised air humidity range for work
Certification
120 a 235 l/min – hood regime, in 5 eligible rates
TFT display
1 420 g
55 to 61 dB
500 to 700 recharging cycles
3 - 4 hours
75 to 130 cm at waistline
0°to 60°C
20 to 95 % Rh
EN 12 941/A2 TH2,TH3 ; EN 12 942/A2 TM3
Estimated operating time
(hours)
Air-flow
(l/min)
Filter type
Particle
P R SL
A2 P R SL
Combined
ABEK2P R SL
Combined ABE1
P R SL
CA Chemical 3F plus
160
10
9,5
9,5
9,5
210
8,5
6,5
6,25
6,5
Declaration of Conformity is available at: https://www.clean-air.cz/doc
Notified body for testing CE:
Occupations Safety Research Institute, v.v.i.
Jeruzalémská 1283/9, 110 00 Prague 1
Notified body 1024
11. List of parts:
CA CHEMICAL 3F
Order code:
Description:
51 00 41
Decontaminable belt
51 00 42
Decontaminable harness
51 00 30EUR
Battery charger
51 00 30UK
Battery charger UK plug
52 00 10
Battery Li-Ion 14,4 V / 2,6 Ah
52 00 43.1
Comfort padded belt 3F
52 00 44.1
Comfort padded harness
52 00 46
Set of decontamination plugs - 3F
71 00 60
Light flexi hose QuickLOCK™ - CA40x1/7“
71 00 86
Rubber hose QuickLOCK™ – CA40×1/7“