10. Electrical specifications
10.1 AC Voltage
Range
Resolution
Accuracy
Imput
Impedance
Fixed
Value Input
45–1000 Hz
> 1–3 kHz
60 mV
0.01 mV
±(1.2 % + 5)
±(2.0 % + 5)
Around
> 3000 MΩ
1000 V DC
or 750 V AC
600 mV
0.1 mV
6 V
0.001 V
±(1.0 % + 3)
±(1.5 % + 5)
Around
10 MΩ
60 V
0.01 V
600 V
0.1 V
750 V
1 V
±(1.2 % + 5)
±(3.0 % + 5)
10.2 DC Voltage
Range
Resolution
Accuracy
Imput
Impedance
Fixed
Value Input
60 mV
0.01 mV
±(0.8 % + 3)
Around
> 3000 MΩ
1000 V DC
or 750 V AC
600 mV
0.1 mV
6 V
0.001 V
±(0.5 % + 1)
Around
10 MΩ
60 V
0.01 V
600 V
0.1 V
1000 V
1 V
±(1.0 % + 3)
10.3 DC Current
Range
Resolution
Accuracy
Overload protection
600 µA
0.1 µA
±(1.0 % + 3)
Fuse 1: F1A H 240 V (CE),
ø 6 × 25 mm
6000 µA
1 µA
60 mA
0.01 mA
600 mA
0.1 mA
6 A
1 mA
±(1.2 % + 5)
Fuse 2: F10A H 240 V (CE),
ø 6 × 25 mm
10 A
0.01 A
Summary of Contents for ut61d