No.
Description
Cause
How to clear
100
Detector not
connected
No detector is connected or the detector
has a failure, such as a broken wire
This is automatically
cleared when a normal
detector is detected.
110
Detector type
error
An unusable detector has been
connected and started.
Connect a usable
derector.
120
Overstroke
The maximum stroke has been
exceeded, due to an upthrust, for
example.
Use the [RESET] key or a
communication command
to make the unit execute
an error reset.
130
Detector
communication
error
There is an error in communication with
the detector, or the detector has a failure
such as a broken wire.
(Same as above)
No.
Description
Cause
How to clear
200
Parameter R/W
error
A fault has occurred during reading or
writing of parameters.
Press the [RESET] key for
at least 5 seconds or use
a communication
command to make the unit
execute an error reset.
210
Internal memory
overflow
The internal memory has overflowed.
Use the [RESET] key or a
communication command
to make the unit execute
an error reset.
No.
Description
Cause
How to clear
300
External
communication
error
One of the following RS-232C
communication errors has occurred:
parity error, framing error, overrun error,
1-second communication stoppage,
10-second communication interruption
due to RTS OFF, or other reason.
Use the [RESET] key or a
communication command
to make the unit execute
an error reset, or transmit
a normal command.
301
External
command error
Communication command format
malfunction
Data exceeds the specified range when
parameter writing is activated by a
communication command.
(Same as above)
310
Limit value
setting
Abnormal relation between upper and
lower limit values of a SET No. parameter
[RESET] key Input again
* If you do not input
revised values,
measurement will be
possible but you will not
be able to obtain correct
judgment results.
311
Parameter
setting error
Data exceeds the specified range when
setting parameters by using the keys.
[RESET] key Input again
320
Position overflow The measurement value has exceeded
the indication range
Adjust the measurement
value to within the
indication range.