5-3-7 Averaging Measured Values
This function stabilizes the display by obtaining moving averates from internal measured
values that are refreshed with a frequency of 1 ms.
Set the count for averaging measured values by setting initial parameter P51.
Note that the response time for signal output varies according to the averaging count.
(With a count of 10, it is 10 ms.)
5-3-8 Detector Constant (Lever Ratio Calculation)
This function multiplies the measurement value by a set value. It enables automatic
conversion of the lever ratio in situations such as measurement through a lever.
Set a detector constant by setting initial parameter P02.
5-3-9 Setting the Display Resolution
The display resolution can be specified by setting initial parameter P03.
Data is displayed or output via RS-232C as follows:
Display:
Displays only the specified digits and hides any rounded up digits.
<Example> When the resolution is 10
µ
m, 1.2345 (mm) is displayed as 1.23 (mm).
RS-232C: Outputs a value with a fixed number of digits, displaying “0” in place of any
rounded up digits.
<Example> When the resolution is 10
µ
m, 1.2345 (mm) is displayed as 1.2300 (mm).
5-3-10 Input Monitor
The input monitor function displays the input states of various signals. You can check the
wiring and signal states in a simple operation. Use it in the event of trouble.
For details on the display, refer to 5-2 “Overall Flow of Operations”.
Lever ratio calculation [A
×
Constant]
Chapter
5
Mode of Use