26
GETEST
Instructions Manual
(Continued) Environmental features
Protection degree
IP20
Mechanical features
Terminal
Power Supply
IEC320 C20 16A
Injection Current
4 mm safety banana CAT IV 600 V 32 A
Voltage measurement
4 mm safety banana CAT IV 600 V 32 A
Dimensions
Weight
45 kg (Electrodes not included)
Enclosure
Steel / aluminium
Despite the fact that the connectors are designed for nominal currents of up to 30 A,
they support currents of up to 150 A for 20 ms without affecting their durability or
characteristics.
A low impedance power line is required to get maximum performance, since peaks of
up to 130 A for 20 ms can appear when maximum power is requested. Wiring with a
section of 2.5mm
2
or greater is recommended to maintain low resistance in the wiring.
Internal protections
Magnetothermal
20 A, curve D EN60989
Internal fuse
F250 mA
Standards
Safety requirements for electrical equipment for measurement, control, and laboratory use�
Part 1: General requirements�
EN 61010-1
Safety requirements for electrical equipment for measurement, control, and laboratory use�
Part 2-030: Particular requirements for testing and measurement circuits�
EN 61010-2-030
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part
1: General requirements
EN 61326-1: 2013
285 mm
285 mm
340 mm
Figure 17: GETEST dimensions�