MAINTENANCE
BLANKETROL III, Model 233
OPERATION AND TECHNICAL MANUAL
Page 78 of 131
K.
After all the water has drained from the unit, disconnect the drain hose, wipe, clean
and store in a dry environment without disinfecting. (Refer to Section (7.)
specifications for storage conditions)
L.
Discard the water drained from the unit.
M.
If any other maintenance/repair is to be completed, go to the appropriate section,
e.g., cleaning the water filter. If not, go to Section (4-2.2.) to replenish the reservoir.
4-2.2. Replenishing the Reservoir
A.
Check that the drain hose is disconnected.
B.
Lift the water fill opening lid and gradually pour approximately 2 gallons (7.6 liters) of
distilled water (or the distilled water-bacteriostatic preparation described in Section
(4-2.) into the reservoir (DO NOT USE ALCOHOL, TAP WATER, OR DE-IONIZED
WATER).
C.
Stop pouring when the water reaches the strainer visible at the bottom of the water
fill opening.
D.
Proceed with normal operations; always check the water level before starting.
4-3.
MAINTENANCE OF THE WATER FILTER
The BLANKETROL III circulating system includes a water filter designed to clear the line of any
particulate matter as the water is pumped through the system. The water filter assembly is a plastic
T-shaped fitting that intersects the hose from the water flow indicator to the flow switch. A stainless
steel wire-mesh screen is located inside the clear plastic cap of the water filter assembly.
Quarterly, or more often if deemed necessary, the water filter should be disassembled and cleaned.
To do so, the reservoir must first be drained and the front storage drawer extended. The cap of the
water filter assembly is unscrewed. The wire mesh and the plastic cap are cleaned. The parts are
then reassembled.
CAUTION
Use distilled water only. Do Not Use De-Ionized water. De-Ionized water
may
cause corrosion to plumbing system components.
Do Not Use Tap Water.
Minerals and deposits
can clog plumbing system components.
Do not
use alcohol. Alcohol may cause blanket deterioration.
Do not
overfill. Overfilling may result in overflow when the water in the blanket
drains back into the system when the system is turned off.
CAUTION
Always drain the BLANKETROL III to a sanitary drain because
bio-contaminants
may be present in the unit’s water supply.