MAINTENANCE
BLANKETROL III, Model 233
OPERATION AND TECHNICAL MANUAL
Page 76 of 131
10.
If unit is to be
returned to service
, replenish reservoir(s) with the appropriate volume
of distilled water.
11.
If
unit is being
placed in dry storage
, continue with procedure.
12.
When all fluid has been removed from the unit, disconnect the drain hose(s) and wipe
unit clean.
13.
Unit is now ready for storage.
14.
Document unit maintenance.
FLUID CIRCUIT DISINFECTION/DRY STORAGE PROCEDURE
for Circulating Water Units Utilizing CE Approved Maranon H (Product of Ecolab)
(Contact your local Cleaning Supplier)
(Maranon H Is Not Available In the United States)
The following procedure disinfects the fluid circuit in these products. The required tools/supplies
are appropriate drain hose(s) and/or hose connector(s), Maranon H, distilled water, and the
appropriate AC electrical power
PROCEDURE:
1.
Drain the water from the reservoir as described in Section 4-2.1.
2.
Turn unit
off (“O” position)
.
3.
Fill the reservoir with the appropriate amount of Maranon H per the below chart.
4.
Fill the reservoir with the appropriate amount of water per the above chart.
5.
Turn the unit on and circulate per the Chemical Cleaning Circulation (CCC) Chart
(above).
Note: The duration indicated in the CCC chart is meant to begin when the
circulating water reaches the temperature indicated in the CCC chart.
6.
Drain the unit as instructed in Step #1.
7.
Rinse the unit with distilled water three (3) times as described in Steps 4-6.
Be sure to
omit the Maranon H.
8.
After the third rinse, drain the unit and fill the reservoir with distilled water and circulate.
9.
Check the water with PH strips or other appropriate test method for Maranon H
detection. If Maranon H is detected, repeat Steps 4-6.
10.
If unit is to be
returned to service
, replenish reservoir(s) with the appropriate volume
of distilled water.
11.
If
unit is being
placed in dry storage
, continue with procedure.
12.
When all fluid has been removed from the unit, disconnect the drain hose(s) and wipe
unit clean.
13.
Unit is now ready for storage.
14.
Document unit maintenance.
UNIT
MARANON H
WATER
BLANKETROL III
227 Milliliters
7.6 Liters