OPERATION
BLANKETROL III, Model 233
OPERATION AND TECHNICAL MANUAL
Page 34 of 131
The above depends upon the relationship of the patient temperature to the
Set Point temperature. Note that the Set Point may be displayed in
Fahrenheit.
5.
The pump is activated.
6.
The heater or compressor may be activated.
7.
The water flow indicator on the right side panel begins to move. The water
moves from the unit through the blanket and returns to the unit.
8.
System operates in GRADIENT 10C mode.
Y.
Press TEMP SET button
1.
The microprocessor board beeps.
2.
The LED in the corner of the button illuminates.
3.
The Status Display returns to set temperature mode.
Z.
Press GRADIENT 10C button and then SMART button
1.
The microprocessor board beeps.
2.
The LED’s in the corner of the buttons illuminate.
3.
The WATER display shows the temperature of the water in the reservoir.
4.
The Status Display shows:
* XXXXXX PATIENT
AUTO SETPT 37.0 C
Or Status Display shows:
PATIENT @SETPT
AUTO SETPT 37.0 C
* (”XXXXXX” represents “HEATING” or “COOLING”.)
The above depends upon the relationship of the patient temperature to the
Set Point temperature. Note that the Set Point may be displayed in
Fahrenheit.
5.
The pump is activated.
6.
The heater or compressor may be activated.
7.
The water flow indicator on the right side panel begins to move. The water
moves from the unit through the blanket and returns to the unit.
8.
Wait 30 minutes for SMART MODE to occur.
9.
System operates in GRADIENT 10C SMART mode.
AA.
Press TEMP SET button
1.
The microprocessor board beeps.
2.
The LED in the corner of the button illuminates.
3.
The Status Display returns to set temperature mode.
AB.
Press the GRADIENT VARIABLE button.
1.
The microprocessor board beeps.
2.
The LED in the corner of the button illuminates.
3.
The Status Display shows:
VARIABLE GRADIENT
OFFSET 0 C