OPERATION
BLANKETROL III, Model 233
OPERATION AND TECHNICAL MANUAL
Page 54 of 131
1.
The switch illuminates green.
2.
The microprocessor board goes through self-test.
3.
The Status Display flashes CHECK SETPT.
D.
Consult the physician's orders to determine the desired patient Set Point
temperature. As a safety precaution, the desired patient Set Point can only be set
between 30°C - 40°C (86°F - 104°F) to operate in
GRADIENT 10C SMART MODE
.
E.
Use the C/F button to choose the desired temperature scale (English membrane
only).
F.
Press the TEMP SET button.
1.
The microprocessor board beeps.
2.
The LED in the corner of the button illuminates.
3.
The Status Display shows a temperature Set Point.
G.
Press the Up arrow or Down arrow to change the Set Point temperature on the
Status Display to the desired patient temperature. The display can only be set
between 30°C - 40°C (86°F - 104°F).
1.
The microprocessor board beeps.
2.
The Set Point temperature in the Status Display changes.
H.
Press the GRADIENT 10C button.
1.
The microprocessor board beeps.
2.
The LED in the corner of the button illuminates.
3.
The PATIENT display shows the patient's actual temperature.
4.
The WATER display shows the actual temperature of the water in the
BLANKETROL III equipment.
5.
The Status Display shows:
* XXXXXX PATIENT
AUTO SETPT 37.0 C
Or Status Display shows:
PATIENT @SETPT
AUTO SETPT 37.0 C
* (”XXXXXX” represents “HEATING” or “COOLING”.)
The above depends upon the relationship of the patient temperature to the
Set Point temperature. Note that the Set Point may be displayed in
Fahrenheit.
6.
The pump is activated. The heater or compressor may also be activated.