OPERATION
BLANKETROL III, Model 233
OPERATION AND TECHNICAL MANUAL
Page 60 of 131
I.
Press the Up arrow or Down arrow to change the gradient variable offset to the
desired value.
1.
The microprocessor board beeps.
2.
The Set Point temperature in the Status Display changes.
J.
Press the GRADIENT VARIABLE button.
1.
The microprocessor board beeps.
2.
The LED in the corner of the switch illuminates.
3.
The PATIENT display shows the patient's actual temperature.
4.
The WATER display shows the actual temperature of the water in the
BLANKETROL III equipment.
5.
The Status Display shows:
* XXXXXX PATIENT
AUTO SETPT 37.0 C
Or Status Display shows:
PATIENT @SETPT
AUTO SETPT 37.0 C
* (”XXXXXX” represents “HEATING” or “COOLING”.)
The above depends upon the relationship of the patient temperature to the
Set Point temperature. Note that the Set Point may be displayed in
Fahrenheit.
6.
The pump is activated. The heater or compressor may also be activated.
7.
The Water Flow indicator on the right side panel begins to move.
8.
The water moves from the unit to the blanket and returns to the unit.
K.
Press the SMART key.
1.
The microprocessor board beeps.
2.
The LED on the SMART button illuminates.
L.
Check the water flow indicator to confirm that the water is circulating.
M.
Touch the hyper-hypothermia blanket to confirm that the blanket is heating/cooling.
N.
To make any changes in the control settings, press the TEMP SET button and begin
again.