HIPOT Tester 19071/19072/19073 User’s Manual
5-8
AC mode
Name
Size (byte)
Unit
Range
Description
Step index
1
-
1~10
Must less or equal original step 1
Mode
1
-
1
AC mode
Source
2
V
0, 50~5000
0:OFF
Ramp Time
2
100mS
0~9990
0:OFF
Reserved
2
-
0
Reserved
Test Time
2
100mS
0~9990
0:Continue
Fall Time
2
100mS
0~9990
0:OFF
High Limit
4
100nA
10~200000
10~30000
The max. is 200000 when EN50191 is OFF.
The max. is 30000 when EN50191 is ON.
Low Limit
4
100nA
0,
10~200000
10~30000
0:OFF
The max. is 200000 when EN50191 is OFF.
The max. is 30000 when EN50191 is ON.
Arc Limit
4
100nA
0, 10000~200000
0:OFF
Reserved
4
-
0
Reserved
DC Mode
Name
Size (byte)
Unit
Range
Description
Step index
1
-
1~10
Must less or equal original step 1
Mode
1
-
2
DC mode
Source
2
V
0, 50~6000
Ramp Time
2
100mS
0~9990
0:OFF
Dwell Time
2
100mS
0~9990
0:OFF
Test Time
2
100mS
0~9990
0:Continue
Fall Time
2
100mS
0~9990
0:OFF
High Limit
4
100nA
1~50000
Low Limit
4
100nA
0~50000
0:OFF
Arc Limit
4
100nA
0, 10000~50000
0:OFF
Inrush
4
-
0/10000
0:OFF, 10000:ON
IR Mode
Name
Size (byte)
Unit
Range
Description
Step index
1
-
1~10
Must less or equal original step 1
Mode
1
-
3
IR mode
Source
2
V
0, 50~1000
Ramp Time
2
100mS
0~9990
0:OFF
Dwell Time
2
100mS
0~9990
0:OFF
Test Time
2
100mS
0, 3~9990
0:Continue
Fall Time
2
100mS
0~9990
0:OFF
High Limit
4
100kOhm
0~500000
0:OFF
Low Limit
4
100kOhm
1~500000
IR Range
4
-
0~6
0:
300nA, 1:
3uA, 2: 30uA, 3: 300uA,
4: 3mA, 5: 5mA, 6: Auto Range
Reserved
4
-
0
Reserved
GC Mode
Name
Size (byte)
Unit
Range
Description
Step index
1
-
1~10
Must less or equal original step 1
Mode
1
-
4
GC mode
Source
2
100mA
0/1
Reserved
2
-
0
Reserved
Dwell Time
2
100mS
1~10
Reserved
2
-
0
Reserved
Reserved
2
-
0
Reserved
High Limit
4
100mOhm
1~50
Low Limit
4
100mOhm
0~50
0:OFF