HIPOT Tester 19071/19072/19073 User’s Manual
2-12
Range
IR Display Value
Setting Voltage
50V ~ 250V
Setting Voltage
250V ~ 1000V
5mA(3~5mA)
0.10M
Ω
~ 0.10M
Ω
0.10M
Ω
~ 1.00M
Ω
3mA(0.3~3mA)
0.10M
Ω
~ 0.90M
Ω
0.10M
Ω
~ 3.50M
Ω
300uA(30~300uA)
0.10M
Ω
~ 9.00M
Ω
0.10M
Ω
~ 25.00M
Ω
22.0M
Ω
~ 35.0M
Ω
30uA(3~30uA)
0.10M
Ω
~ 25.00M
Ω
22.0M
Ω
~ 90.0M
Ω
0.10M
Ω
~ 25.00M
Ω
22.0M
Ω
~ 250.0M
Ω
0.220G
Ω
~ 0.350G
Ω
3uA(0.3~3uA)
22.0M
Ω
~ 250.0M
Ω
0.220G
Ω
~ 0.900G
Ω
25.0M
Ω
~ 250.0M
Ω
0.220G
Ω
~ 2.500G
Ω
2.20G
Ω
~ 3.33G
Ω
300nA(20~300nA)
0.200G
Ω
~ 2.000G
Ω
0.200G
Ω
~ 2.500G
Ω
2.20G
Ω
~ 50.00G
Ω
Note
Select IR suitable current range please follow test voltage and DUT
insulation resistance for counting the quantity of current then follow it to
select suitable current range.
RAMP: Step-up setting voltage needed time, 0 means OFF.
DWELL: Set the required time for test, 0 means OFF. (Do not judge the high and
low limit for insulation resistance during this period.)
FALL : Set the required time for falling from selected voltage to zero, 0 means OFF.
(4) Ground Continue Mode
STEP 1/1
GC
LOW : OFF
UP
CURR : 0.0A
DOWN
HIGH : 1.0
Ω
DWELL: 0.3s
NEXT
EXIT
SELECT MODE
RMT LOCK OFST ERR
CURR : Set the output current (0.1A maximum).
HIGH : Set the high limit of ground continue (0.1
Ω
∼
5
Ω
)
.
DWELL : Set the time required for test (0.1 sec
∼
1 sec). (Do not judge the high
and low limit for ground continue during the time.)
LOW
: Set the low limit for ground continue with the value lower than the high
limit of ground continue or OFF.
(5) Pause Mode
STEP 1/1
PA
UP
PAUSE:PAUSE-MODE
DOWN
UNDER TEST SIGNAL:OFF
NEXT
EXIT
SELECT MODE
RMT LOCK OFST ERR