HIPOT Tester 19071/19072/19073 User’s Manual
4-4
4.2.4 IRV Calibration (DC/IR Mode)
(1) Press [UP] or [DOWN] to select IRV offset calibration.
(2) Connect a DC high voltage meter to HiPot tester.
(3) Press [STOP][START] to read the high voltage meter value, ex. 0. 052kV.
(4) Press [INC.] or [DEC.] until shows 0. 052 kV and then press [ENTER] for saving.
(5) Press [STOP] to stop IR voltage offset calibration.
(6) Press [UP] to change to the next calibration item.
(7) Press [STOP][START] to read the high voltage meter value, ex. 1.052kV.
(8) Press [INC.] or [DEC.] until shows 1.052 kV and then press [ENTER] for saving.
(9) Press [STOP] to stop IR voltage full scale calibration.