Electrical Safety Analyzer 19032-P Quick Start Guide
42
press
START
on panel or re-trigger START signal on
rear panel.
(2) The setting is 0.3 ~ 999sec: When the setting time is up
then end the pause mode.
Short/Open Circuit Detection Mode (OSC)
OPEN CHK: It sets the judgment test result to open condition(compare the
test reading with the read standard capacitance value [Cs]).
SHORT CHK: It sets the judgment test result to short condition(compare
the test reading with the read standard capacitance value
[Cs]).
CHNL (H-L): The setting is the same as AC/DC CHANNEL.
4.9
CALIBRATION Function
4.9.1 Enter Calibration Method
1. Open the upper cover, press
SW402
and then powered the analyzer
on.
2. When the title bar shows “MAIN MENU”, press numerical key which is
corresponding to
CALIBRATION
then will show “ENTER
CALIBRATION PASSWORD” window.
3. By using numerical keys to input PASSWORD [7] [9] [3] [1].
4. Press
ENTER
, select
[DEVICE]
and then enter calibration procedure.
4.9.2 Clear Memory
1. When title list shows “MAIN MENU”, press numerical key that
corresponds to
CALIBRATION
then it will show “ENTER
CALIBRATION PASSWORD” window.
2. By using numerical keys to input PASSWORD [8] [5] [2] [4] [6].
3. After pressing [ENTER] key, “MESSAGE” window will be appeared.
Users can select if clear memory by Function Keys [YES], [NO] or
press [EXIT] to abort memory clearance.
4. If Function Key [YES] is selected, all of saved data will be cleared, all
setting parameters will be reset as initial value.
5. After clearing the memory, Option parameter needs to be reset.