Electrical Safety Analyzer 19032-P Quick Start Guide
WARNING
Please use correct specification when replace fuse
or may cause hazard.
18. Normal operation of the unit is AC power
If power is unstable, it may cause the unit function is not actual or
abnormal. Therefore, please use appropriate equipment turn to
suitable power such as power stabilizer.
19. Output power is 500VA
When DUT drawing mass current before deadline of fail judgment and
output current, it may flows mass current (about ten amperes) up to ten
milliseconds. Before processing test may be the same condition.
Please notice the capacity of power cord and the current cable of
linking with other instrument or equipment.
20. Storage
The unit normal operation temperature humidity range is 5
°
C
∼
40
°
C,
80% RH. If over this range then function may malfunction. Please
don’t position the equipment so that it is difficult to operate the
disconnecting device. The unit storage temperature range is -10
°
C
∼
50
°
C, 80% RH. If you don’t use it for a long time, please use original
material packing and then store it. For correct test and safety, please
keep it from direct sunlight or high temperature, vibration, humidity and
dusty place.
21. Warm up
All functions of the analyzer are activated when the power switch is
turned on. However, to attain the precision in the specification, please
warm the instrument over 15 minutes.
22. Warning signal of testing
“DA
UNAUTHORIZ
NGER – HIGH VOLTAGE TEST IN PROGRESS,
ED PERSON KEEP AWAY”
23. TWIN PORT
The unit process twin port measurement mode with AC withstand, DC
withstand or IR (insulation impedance) on GB (grounding impedance).
When process the maximum AC or DC output current of this
measurement mode, please don’t work over 1/2 the maximum
specification continuously.
12