Electrical Safety Analyzer 19032-P Quick Start Guide
3
Detection Current
AC: 1mA – 20mA, DC: 1mA – 10mA, resolution
0.1mA
Min. pulse width
40us 20us 10us 4us Approx.
TwinportTM function (SPC)
Functions
WV and GB test can be performed at the same
time. (ON/OFF)
Insulation Resistance Test
(Note 4)
Test Voltage
DC: 0.05kV ~ 1kV, Constant Voltage
V-display Accuracy
± (2% of r 0.5% of full scale)
Resistance Range
0.1M
Ω
∼
50G
Ω
Measuring
Accuracy
≥
500V:
1
∼
1000M
Ω
: ± (5% of r 0.5% of full
scale)
1001
∼
9999M
Ω
: ±(10% of r 0.5% of
full scale)
10G
Ω
∼
50G
Ω
: ±(15% of r 1% of full
scale)
< 500V:
0.1
∼
1000M
Ω
: ±(10% of r 0.5% of full
scale)
< 100V:
0.1
∼
1000M
Ω
: ±(15% of r 0.5% of full
scale)
Test Time(Note 2)
0.3 ~ 999 Sec. Continue
Ground Bond Test (Note 5)
Output Current
3.00 ~ 40.00Aac. Constant Current, 0.01A step
Current Accuracy
± (2% of s 0.1% of full scale)
Output Waveform
50Hz, 60Hz
±
0.1%, sine wave
Current Meter
0.00
∼
40.00A
Meter Accuracy
± (1% of r 0.17% of full scale)
Resistance Range
10.0 ~ 510.0m
Ω
(with offset value)
Resistance
Accuracy
± (2% of r 0.1% of full scale) at
≥
10A
± (2% of r 1% of full scale) at
<10A(<210m
Ω
)
± (3% of r 2% of full scale) at
<10A(>210m
Ω
)