14
11. Quality Assurance Tests
The test result of each step should meet the following passing criteria.
Otherwise, it may be considered as failed test.
♦ Key Function Test
When the functional keys are pressed step by step, the outputs of the device should be identical
to the following description:
1.
Press the ON / OFF button to turn on the device. The LCD screen will show 7 modes,
minutes left, and A&B output channels.
2.
When the MODE key is pressed, the corresponding mode is selected on the LCD screen.
3.
Press the & button, the stimulation intensity is adjusted. The default level is 0.
4.
The number displayed in the top left corner is the remaining stimulation time.
♦ LCD Check
The information displayed on the LCD screen should be identical to the following description:
Output Channels
Battery indicator
Stimulation Timer
Current Stimulation Mode
Level of Intensity
Bluetooth indicator * (optional)
*
Bluetooth is an OPTIONAL function for the device. The Bluetooth indicator flashes if no
bluetooth connection and stays lit when that connection is established.
*
Operation to the device is available through APP when they are connected via Bluetooth.
12. Accessories
● Connecting Wires
The lead wires consist of PVC material in the outside and thin copper wire on the inside. The
total length of the lead wire is 59.05 inches and there are 3 connectors on each wire. The one
connector on one end is used to connect to the device and the 2 connectors on the other end
are used to connect to the electrode pads. The pulse generated by the device to the pads
through these connecting wires.
● Electrode Pads
The electrode pads are reusable up to 80 uses.