102
-
Do not disconnect the power supply when the data is being accessed.
Take care as the contents of the CF card may get destroyed and the data may get lost,
when the power supply of KR2000 is disconnected, when the data file is being accessed.
Do not remove the USB cable when the data is being accessed.
Operations of KR2000 system may get unstable on removing the USB cable, when the
data file is being accessed. Moreover, in case to remove the cable, click on “Safe
Removal of Hardware” icon displayed on the task tray on the PC screen and then
execute the operations stipulated by OS.
Do not insert and remove the USB cable from the instrument during recording.
Performance of the instrument will become unstable.
Access to the data when it is being recorded
KR2000 system executes the access to the CF card when the storing operation is in
progress. Although it can operate exclusively by the access from the PC, however, as the
total performance decreases, access the data when the storing operation is being
stopped,
Update card information
After connecting the cable, the file on the updated CF card may not be displayed on the
PC. When recording operation, write of setup file, snapshot is executed after connecting
the cable, file information is updated by inserting and removing the CF card.
Warning