BEFORE YOU START

The reactor is now ready for use. Before filling the reactor, make sure the 0-14mm drain valve is in the closed position. Once 

setup is complete, the reactor can be moved into position and connected to the circulator. Keep the wheels locked for 

added safety. 

 

Chemglass presumes some knowledge of this type of equipment on the part of the end user. Properties such as mechanical 

strength of glass, thermal stress introduced to the reactor from exothermic reactions, allowable temperature differentials, 

pressure and vacuum must all be considered with extreme caution. 

If you have any questions, please contact our technical service department at (800) 843-1794 or 

[email protected]

 

Complete Assembly Instructions 

The jacketed process filter reactor, the fully assembled reactor head, stirrer shaft & agitators, Tru-Stir™ stirrer shaft coupling, 

mechanical seal stirrer bearing, temperature probe & adapter, and motor are shipped assembled on the support frame. Use 

the following instructions should you need to remove or adjust these components. 

 

1.

 

Remove torque knobs from reactor support clamp and remove PTFE coated support clamp. 

2.

 

Please refer to filter base assembly then place filter base in the bottom of 

the reactor stand, positioning detachable drain valve toward the front of 

the stand. 

3.

 

Carefully place the reaction vessel on the filter base and line up with the 

vessel. Replace the front reaction vessel clamp and tighten the torque 

knobs evenly until you hear a “click”. 

4.

 

Secure the vessel and filter base together with the CG-141-T quick release 

PTFE coated clamp, making sure that the entire o-ring is seated in the 

groove. Adjust the threaded bolt to improve the seal around the flange. 

5.

 

Attach the S.S. beaded pipe adapters (CG-1969-A) to the inlet/outlet of 

the reaction vessel via the 1” beaded pipe coupling (CG-1968). Tighten the 

coupling to specified torque setting.  

6.

 

Place the 200mm Perfluoro o-ring into the groove of the reaction vessel 

flange. 

7.

 

Assemble the stirrer shaft and agitators and lower into the reaction vessel. 

8.

 

While lowering the reaction vessel lid onto the vessel, carefully insert the 

end of the stirrer shaft through the 45/50 center neck of the lid. PLEASE NOTE: THE MECHANICAL SEAL STIRRER 

BEARING IS NOT INSERTED IN THE HEAD AT THIS TIME. 

9.

 

Dismantle then slide the mechanical seal stirrer bearing over the end of the stirrer shaft and “press “into the 

center neck, sealing the o-ring. 

10.

 

Position the powder fill port towards the front of the reactor. 

11.

 

Secure the vessel and lid together with the quick release PTFE coated clamp, making sure the entire o-ring is 

seated in the groove. Adjust the threaded bolt to improve the seal around the flange. 

12.

 

Attach the Tru-Stir stirrer shaft coupling to the 1/4 HP electric stirrer motor shaft. Then attach the motor to the 

support frame via the supplied four socket head cap screws. 

13.

 

Insert the end of the stirrer shaft into the stirrer shaft coupling and tighten the collar with an Allen wrench. 

Check the vertical alignment of the reactor and adjust if necessary. 

14.

 

Tighten the aluminum nut on the mechanical seal stirrer bearing, which compresses the bearing onto the 

stirrer shaft. 

15.

 

Attach the motor controller/temperature monitor to the support frame via the small Kwik Klamp II and 1/2" 

OD 90° S.S. support rod. 

16.

 

Attach the control cord from the electric stirrer motor to the bottom panel of the controller. Make sure that 

the speed control knob on the controller is turned completely off BEFORE turning the controller on. If using the 

air motor, attach air source to the motor. The brass needle valve on the back of the air motor should be 

CLOSED. 

17.

 

Loosen compression nut and insert the thermocouple thru the thermocouple adapter and then insert the 

adapter into the desired side neck. Adjust the depth of the thermocouple by tightening the Chem-Thread at 

the top of the adapter. PLEASE NOTE: CHECK TO SEE THAT THE THERMOCOUPLE DOES NOT COME INTO 

CONTACT WITH EITHER OF THE AGITATORS BY MANUALLY TURNING THE STIRRER SHAFT. Attach the thermocouple 

cord to the probe then to the temperature monitor. 

18.

 

Attach the condenser and any other peripheral glassware. 

*NOTE: THE BOTTOM IMPELLER SHOULD NOT BE RUBBING ON THE FILTER*

 

 

Summary of Contents for JACKETED FILTER

Page 1: ...IONS FOR JACKETED FILTER REACTOR SYSTEMS Support Frame with Oversized Rugged S S Uprights and Heavy Duty PTFE Coated Aluminum Fittings Mechanical Seal Stir Bearing PTFE Filter Base Open Frame Design f...

Page 2: ...rates in case you need to return items for repair or replacement Crate 1 Unpack the support frame with reactor and components by removing the plastic electrical ties that hold the reactor in place and...

Page 3: ...Klamp II 1ea CG 3498 03 90 Support Rod 1 The support frame has a universal motor mount It can be used with an vertical or horizontal electric air or optional explosion proof XP motor The mount automa...

Page 4: ...embly CG 2095 slides over the end of the shaft The height from the bottom will depend on the total volume you intend to run in the reactor 3 Once you have the upper agitator in place tighten the compr...

Page 5: ...it is pushed down on all edges evenly 5 Wrap the lower outlet adapters with NPT end with PTFE tape Thread the adapter into the filter base until it becomes snug DO NOT OVER TIGHTEN 6 Attach the beade...

Page 6: ...1 glass beaded pipe on the inlet outlet 2 Loosen the nut on the beaded pipe coupling Wet the PTFE Viton liner inside the coupling and attach one side to the 1 beaded pipe on the jacket 3 Insert the b...

Page 7: ...Assemble the stirrer shaft and agitators and lower into the reaction vessel 8 While lowering the reaction vessel lid onto the vessel carefully insert the end of the stirrer shaft through the 45 50 ce...

Page 8: ...www cglifesciences com technical service cglifesciences com...

Page 9: ...www cglifesciences com technical service cglifesciences com Optional Components Safety Shield Side Support Shelf Distillation Side Support Shelf Powder Addition Funnel Data Logger...

Page 10: ...mA or 10 V Zener barriers contain intrinsically safe circuits that are to be used to drive intrinsically safe field devices with hazardous area The manufacturer s data sheets must be consulted The re...

Reviews: