background image

www.cglifesciences.com 

 

 

[email protected] 

Overhead Stirrer Motor 

   

 

                                      Air Motor                                    1/4 HP Vertical Motor 

 

 

 

                             1/4 HP Horizontal Motor                     Explosion Proof (XP) Motor 

 

The motor is installed on the support frame when shipped. Use the following instructions should you 

need to remove or adjust these components. 

 

Components Needed for Overhead Stirrer Motor: 

1ea  CG-2033-B-25  1/4 HP Vertical Electric Stirrer Motor 

1ea  CG-2025-20 

Air Motor* (Optional) 

1ea  CG-2033-V-75  Motor Controller/Temperature Monitor (Only Supplied with 1/4 HP Electric  

 

 

 

Stirrer Motor) 

1ea  CG-2044 

Tru-Stir™ Stirrer Shaft Coupling 

1ea  CG-9253-10 

Small Kwik Klamp II 

1ea  CG-3498-03 

90° Support Rod 

 

1.

 

The support frame has a universal motor mount. It can be used with an (vertical or horizontal) 

electric, air, or optional explosion proof (XP) motor. The mount automatically centers the motor 

directly above the reactor. 

2.

 

The 1/4 HP electric motor is shipped attached to the support frame via four socket head cap 

screws.  

3.

 

When using the 1/4 HP electric motor, the motor controller/temperature monitor, with 

mounting bracket on side panel, needs to be mounted on the S.S. upright. Use the small Kwik 

Klamp II and the 90°S.S. support rod to mount the controller to the support frame. Tighten all 

knobs securely.  

 

*The Air Motor requires the air supply be filtered and a lubricator be installed between the air source 

and motor. Filter-Regulator-Lubricator (CG-2025-10) is available separately. 

Summary of Contents for JACKETED FILTER

Page 1: ...IONS FOR JACKETED FILTER REACTOR SYSTEMS Support Frame with Oversized Rugged S S Uprights and Heavy Duty PTFE Coated Aluminum Fittings Mechanical Seal Stir Bearing PTFE Filter Base Open Frame Design f...

Page 2: ...rates in case you need to return items for repair or replacement Crate 1 Unpack the support frame with reactor and components by removing the plastic electrical ties that hold the reactor in place and...

Page 3: ...Klamp II 1ea CG 3498 03 90 Support Rod 1 The support frame has a universal motor mount It can be used with an vertical or horizontal electric air or optional explosion proof XP motor The mount automa...

Page 4: ...embly CG 2095 slides over the end of the shaft The height from the bottom will depend on the total volume you intend to run in the reactor 3 Once you have the upper agitator in place tighten the compr...

Page 5: ...it is pushed down on all edges evenly 5 Wrap the lower outlet adapters with NPT end with PTFE tape Thread the adapter into the filter base until it becomes snug DO NOT OVER TIGHTEN 6 Attach the beade...

Page 6: ...1 glass beaded pipe on the inlet outlet 2 Loosen the nut on the beaded pipe coupling Wet the PTFE Viton liner inside the coupling and attach one side to the 1 beaded pipe on the jacket 3 Insert the b...

Page 7: ...Assemble the stirrer shaft and agitators and lower into the reaction vessel 8 While lowering the reaction vessel lid onto the vessel carefully insert the end of the stirrer shaft through the 45 50 ce...

Page 8: ...www cglifesciences com technical service cglifesciences com...

Page 9: ...www cglifesciences com technical service cglifesciences com Optional Components Safety Shield Side Support Shelf Distillation Side Support Shelf Powder Addition Funnel Data Logger...

Page 10: ...mA or 10 V Zener barriers contain intrinsically safe circuits that are to be used to drive intrinsically safe field devices with hazardous area The manufacturer s data sheets must be consulted The re...

Reviews: