TOTAL HARMONIC DISTORTION
THD%
1 range only
Measurement extent
0.5% to 600.0%
Resolution
0.1%
Triggering threshold
- on voltage from 300mV
- on current from 300mA
Below the triggering threshold, display of 4 hyphens and
Accuracy
THD%
Specified measu-
0.5% to 99.9%
100.0% to 200.0%
rement range
Accuracy
10% R + 5 counts
15% R
NB :
Base frequency : For triangular and square signals, with a 50% cyclical ratio, at 50Hz
± 1Hz and at 60Hz ± 1Hz, the accuracy in % of the reading (R) is 3% R + 2 counts.
Frequency range
The input signal must be stationary and the fundamental between 45Hz and 65Hz.
Outside this frequency range or if the signal is not stationary, display of at the
same time as the displayed value, to indicate that the measurement is outside the
specified range.
Response time of the digital display : one value per second.
Permitted overload and warning signal : idem Currents and Voltages.
DISTORTION FACTOR
DF%
1 range only
Measurement extent
0.5% to 100.0%
Resolution
0.1%
Accuracy
10% R + 5 counts
43
44
NB :
Base accuracy : For triangular and square signals, with a 50% cyclical ratio, at 50Hz
± 1Hz and at 60Hz ± 1Hz, the accuracy in % of the reading (R) is 3% R + 2 counts.
Triggering threshold, frequency range and response time of the display : idem THD%.
Permitted overload and warning signal : idem Currents and Voltages.
CREST FACTOR
CF
1 range only
Measurement extent
1.00 to 10.00
Resolution
0.01
Accuracy
10% R + 3 counts
Triggering threshold : idem THD%.
Frequency range : from 40Hz to 450Hz.
Response time of the digital display : one value every 500ms.
Permitted overload and warning signal : idem Currents and Voltages.