10.3.2 Measurement input protection
CAUTION:
The instrument contains high rupturing capacity fuses (M - 3.15 A - 500 V - 10 kA -
6.3 x 32 mm) with a trip capacity of 10 kA which do not come into effect in correct
use. If the fuses blow, “- E -” is displayed. In this case, inspection of the C.A 6115N
by a recognised service centre is necessary.
10.4 Storage
If the C.A 6115N remains in storage or unused for a prolonged period of time, charge the batteries from
time to time to avoid any damage.
10.5 Metrological check
IIt is essential that all measuring instruments are regularly calibrated.
We advise you to check this instrument at least once a year. For checking and calibration of your instrument,
please contact our accredited laboratories (list on request) or the Chauvin Arnoux subsidiary or Agent in
your country.
10.6 After-sales service
Repairs under or out of guarantee: please return the product to your distributor