Intelect Legend XT® Therapy System
72
INSTALLATION/REMOVAL
Remove the Therapy System from the
Therapy System Cart, if equipped.
Refer
to pages 78 through 79
for proper
instructions.
Place Therapy System face down on a clean
working surface protected with a soft, clean
fabric to prevent damage to the lens.
REMOVING THE CHANNEL 3/4 ELECTROTHERAPY AND NiMH BATTERY MODULE
Disconnecting Mains Power
DISCONNECT THE SYSTEM FROM THE POWER
SOURCE (OUTLET OR REMOVE BATTERY
MODULE IF INSTALLED) BEFORE ATTEMPTING
ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES
TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM.
Disconnect the Mains Power Cord from
the power supply. Remove the Rear Access
Panel and disconnect the Mains Power
Cord from the Therapy System.
REAR
ACCESS
PANEL
MAINS
POWER
CORD
Removing Therapy System from Cart
Remove the Front Access Cover and
disconnect the Lead Wires and Accessories
from the Therapy System and Channel 3/4
Electrotherapy Module.
Removing Lead Wires and Accessories
REMOVE LEAD
WIRES AND
ACCESSORIES
NOTE:
Keep Lead Wires and accessories
for later re-installation to the
Therapy System.