INSTALLING THE CHANNEL 3/4 ELECTROTHERAPY AND NiMH BATTERY MODULE (CONTINUED)
Securing Therapy System to Module
Carefully place the Therapy System and
Module on one side. With a #1 Phillips
Screwdriver, install the four 4 mm x 20 mm
screws.
Tighten screws until the Module does not
move on the Therapy System.
Attaching the Lanyard
With a #1 Phillips Screwdriver, remove the
screw retaining the existing Front Access
Panel.
Install Lanyard to the new Extended Front
Access Panel using the same screw.
Do not twist Ribbon Cable. If power is
applied to the system with misalignment
of pins or a twisted ribbon cable, the
controlling electronics in the Module will
be destroyed and possible damage to the
System's internal components could occur.
Setting Therapy System onto Module
Set the Therapy System on the Module.
Make certain the Feet of the Therapy
System are resting in the Module Indents.