2) Set start frequency and stop frequency
Fig. 3.6 Set Start Frequency and End Frequency
3) Set center frequency and frequency span
Fig. 3.7 Set Center Frequency and Frequency Span
The
【
Freq
】
shortcut key in the functional key area of the front panel can be used to set frequency
quickly.
3.4 Signal Power Level Setup
The power level refers to the power level of the source output signal of the test port of the analyzer.
The port of the 3672 series vector network analyzer can set following power level indicators:
Table 3.1 Power Level Index of the Source Output Signal
Frequency range
Source power range (dBm)
Configuration conditions
10 MHz ~ 13.5 GHz/26.5 GHz/40 GHz/50
GHz/67 GHz
-30 ~ +20
Standard configuration
10 MHz ~ 13.5 GHz/26.5 GHz
-100 ~ +20
With programmable step
attenuator options
10 MHz ~ 40 GHz/50 GHz/67 GHz
-90 ~ +20
With programmable step
attenuator options
Select the menu path:
[Stimulus]→[Power]
.
In addition,
【
Power
】
shortcut key is in the
functional key area and shortcut menu bar of the front panel.
Enter the target stimulus power value
in the port power input box, and then click OK.