Aridus II™ Operator’s Manual
Chapter 6: Maintaining the Aridus II™
39
Figure 6-9.
Detachment of Multi-Wire Connector to Membrane Desolvator.
Cleaning the Membrane Desolvator
The following procedure is recommended to clean the membrane desolvator.
1
Locate the membrane rinse kit in the Aridus II
™
completion kit. Using the
adaptors and tubing provided in the rinse kit, connect the two ends from the
“Y” union to the SWEEP GAS OUT port and the port of the upper block of the
membrane desolvator. The open ends of the tubing used in the rinse kit are
labelled. See Figure 6-10 for connections.
Figure 6-10.
Setup of Rinse Kit for Cleaning Membrane Desolvator Assembly.