Aridus II™ Operator’s Manual
20
4
Connect the other end of the nebulizer gas line to the side gas port of the
C-Flow PFA nebulizer. Insert the 4 mm tubing into the PFA nut and finger-
tighten the nut. See Figure 3-8.
Figure 3-8.
Attachment of Nebulizer Gas Line to Gas Port of C-Flow PFA
Nebulizer.
5
Insert the C-Flow PFA nebulizer into the spray chamber port. See Figure 3-9.
Ensure that the nebulizer is fully inserted through both o-rings.
Figure 3-9.
Insertion of C-Flow PFA Nebulizer into Spray Chamber.
6
Carefully set the C-Flow uptake line through the sample inlet notch in the
upper left front of the Aridus II
™
door. See Figure 3-10. Attach the sample probe
assembly to the ASX-112FR probe holder. (Refer to the
ASX-112FR Operator’s
Manual
for further instructions.)