9
2
Preparing for
Installation
Before you install the Aridus II
™
nebulizer system, you should evaluate the
physical arrangement of the laboratory to choose a suitable location. Once you
choose a location, you must carefully unpack the system prior to beginning the
installation.
Choosing a Location
Choosing a location for the Aridus II
™
involves evaluating the laboratory
environment for the availability of space and power. For the Aridus II
™
to
function optimally, the location you select must meet specific requirements
associated with each of these items. The following sections discuss space and
power requirements.
Space Requirements
Most analytical applications benefit from the shortest sample flow path. A one
meter length of Teflon
®
lined PVC tubing is provided to connect the Aridus II
™
to the ICP-MS torch. Place the Aridus II
™
sample out port less than one meter
from the ICP-MS torch. The system without the ASX-112FR autosampler is 34
cm x 25 cm x 56 cm (W x H x D). Adequate space should be allowed for proper
ventilation.
Power Requirements
Place the Aridus II
™
within 1.2 meters of the power outlet. The Aridus II
™
is
provided with a voltage selection switch on the bottom panel, right front. The
voltage may be selected as either 100-120 VAC +/- 10%, 50/60 Hz, 6A or 220-
240 VAC +/- 10%, 50/60 Hz, 3A. See Figure 2-1.
Summary of Contents for Aridus II
Page 1: ...CETAC Aridus II Desolvating Nebulizer System Manual Part Number 480144 rev2 Operator s Manual...
Page 12: ...Aridus II Operator s Manual 12 This page is intentionally blank...
Page 26: ...Aridus II Operator s Manual 26 This page is intentionally blank...
Page 42: ...Aridus II Operator s Manual 42 This page is intentionally blank...
Page 50: ...Aridus II Operator s Manual 50 This page is intentionally blank...
Page 60: ...Aridus II Operator s Manual 60 This page is intentionally blank...
Page 62: ...Manual Part Number 480144 rev2 Printed in USA...