CBX
Calibration
Mode
Setup
: The CBX must be set to this conf. mode before changing
calibration parameters or using “
MANUAL UP/DOWN
” button.
Trig
: This is the normal working mode. In this mode the CBX will
take a sample on every external trig. Set CBX back to Trig for normal
use..
Manual
up/down
Enables the BB1/BB2 to control the sensor in the CBX to go UP and
DOWN. In this mode the depth and concentration are displayed in
the BB2. . Must change “
MODE
” to “
TRIG
” to activate.
Take
sample
No
/
Zero
/
Sample
-‐ Chose the type of calibration. Zero is for clean
water and Sample is sludge sample or concentration.
Cons
Current concentration.
Sample
Enter concentration from Lab test – concentration/suspended solids
value for Sample.
Cleaning
Cleaning
Yes/No
Yes, flush cleaning is switched
ON
. No, the
flush is switched
OFF
.
Manual up/down
Enables the BB1/BB2 to control the sensor in the
CBX to go UP and DOWN. In this mode the sensor
is switched off and can be left out for an unlimited
time for cleaning. Must change “
MODE
” to
“
TRIG
” to activate.
Scale / Alarm
Max
Max level or 20 mA output,
distance E in fig 2
Min
Min level or 4 mA output
Hi-
Alarm
Max level for alarm
Low-
Alarm
Min level for alarm
Alarm
Relay
Pick which alarm relay is to be used on the
BB1/BB2
System
Type
Type of sensor, read only information
Serial
Serial number of the sensor, read only information
SoftW
Software version of the sensor, read only
information
Info
Ch1
Info, measured light value on channel 1
Ch2
Info, measured light value on channel 2
Samples
Counter that shows total number of samples taken
by the CBX from the last power off.