6.4) Change the "pump control mode"
Control mode
Associated pictogram
Function
RADIO ON
default setting
RC ON
Control of the pump by means of radio control RRC1 only.
The control button (11) of the pump is disabled, except
for waking up the pump.
RADIO OFF
RC OFF
Control of the pump by means of the control button (11) of
the pump only.
The radio control receiver and the radio
control RRC1 are disabled
.
To change the "pump control mode", proceed as follows:
- Select screen 5 from the “main menu” (Ref. to
section 6.1
), hold the finger on the touch button (9) until a
confirmation “beep” is heard, the choice made is shown by the related pictogram.
When the battery is removed from the pump or after 1 hour of inactivity (display OFF),
once you wake the pump up, it will always start with the default setting, RADIO ON.
6.5) Alarms / Warning / Notifi cation
These appear on the display during operation or at the end of the cycle, informing the operator
on the state of the pump. It can be combined with beeps.
Message
Meaning
Description
BATTERY
Ï
Ð
BATTERY
LOW
Replace or recharge the battery. NOTE: when the battery volt-
age falls below a minimum safety threshold,
the pump will not
start
, although it is still possible to end a work cycle in progress.
BATTERY
Ï Ð
BATTERY
TEMPERATURE
HIGH
Remove the battery and wait until it cools down. In order to
cool it faster, it is possible to insert it into the battery charger
supplied, thus making use of the specifi c “AIR COOLED” function.
Pump's LED light fl ashes quickly (0.1 s ON 0.1 s OFF) and the
buzzer emits a coordinated audible signal.
max 90°
90°
MOTOR-PUMP
TEMPERATURE
HIGH
The maximum permitted operating temperature of 194 ° F (90
°C) has been reached.
The pump stops
; in this instance wait for
it to cool down. Only when the permitted working temperature
is reached will it be possible to re-use the pump.
INSUFFICIENT
OIL
This appears when the pressure of the hydraulic circuit doesn't
increase but remains near to zero for a duration of 60 consecutive
seconds. Check the oil level and if necessary refi ll (Ref. to 7.4).
30001
3
Ï
Ð
REQUEST
MAINTENANCE
Number of cycle to recommended maintenance interval
reached; the pump continues to work, however, it is recom-
mended that it is sent to
Cembre
for a complete overhaul
(Ref. to section 8).