Q-Dry

 Solvent Evaporator

The Q-Dry offers rapid and efficient evaporation of samples to complete 
dryness. The design of the metal block allows for usage of either the 40 mL or 
60 mL EDGE collection vials. With full control over temperature and nitrogen 
flow, the evaporator can be adjusted to suit any extract. 

 

WARNING

 

Proper precautions must be taken to avoid contact with reagents 
or reagent vapors. Protective gear should be worn as outlined in 
the user’s safety program for hazardous materials and the reagent 
manufacturer’s safety data sheet. Refer to these guidelines for 
proper handling and disposal of the reagents. Dispose of all waste 
in accordance with all applicable local, state, and federal health and 
safety recommendations.

 

HOT

 

Avoid prolonged exposure to metal block positioned on the dry block 
heater. Surface is hot to the touch.

Read and understand the VWR Dry Block Heater Instruction Manual, (P/N 
549040-00) before installing the Q-Dry.

System Installation

1. 

Place the dry block heater in the fume hood.

 

WARNING

 

The dry block heater must be installed in an adequately rated fume 
hood.

2. 

Place the heating block 

A

 onto the dry block heater.

3. 

Place the rod with attached regulator and flow meter 

B

 onto the dry block 

heater.

4. 

Place the adjustable nitrogen manifold 

C

 onto the rod. 

5. 

Connect the supplied 1/4” tube 

D

 from the flow meter to the manifold.

6. 

Connect a 1/4” tube from the nitrogen source to the push-to-connect 
fitting on the regulator. The tube and connections to the nitrogen source 
are not supplied with the Q-Dry instrument. A minimum of 80 psi is recom-
mended.

7. 

Locate the VWR instruction manual (P/N 549040-00) for installation 
instructions regarding the dry block heater.

A

D

C

B